Literature DB >> 1985525

Electrosurgery-induced endotracheal tube ignition during tracheotomy.

A Aly1, M McIlwain, J A Duncavage.   

Abstract

Electrosurgery was the most common source of ignition for operating room fires prior to the advent of lasers. When combined with volatile anesthetic mixtures, electrosurgery has caused ignition of plastic, rubber, paper, enteric gases, and combustible preparation solutions. We report on an intubated patient whose polyvinyl chloride endotracheal tube ignited during a tracheotomy performed with an electrosurgical unit. The oxygen-rich environment, the polyvinyl chloride tube, and the heat generated by the electrosurgical unit combined to produce a fire. Since otolaryngologists are called upon often to perform tracheotomies on intubated patients, it is imperative that they understand the factors involved in the development of such a fire. This case is presented with an explanation of why this type of fire occurs. A brief review of the literature is included. Different kinds of electrosurgical units, precautions as to their use, and the management of electrosurgery-induced endotracheal tube fires are also discussed.

Entities:  

Mesh:

Year:  1991        PMID: 1985525     DOI: 10.1177/000348949110000105

Source DB:  PubMed          Journal:  Ann Otol Rhinol Laryngol        ISSN: 0003-4894            Impact factor:   1.547


  5 in total

Review 1.  Airway fire during tracheostomy: prevention strategies for surgeons and anaesthetists.

Authors:  M L Rogers; R W Nickalls; E T Brackenbury; F D Salama; M G Beattie; A G Perks
Journal:  Ann R Coll Surg Engl       Date:  2001-11       Impact factor: 1.891

Review 2.  Crisis in the operating room: fires, explosions and electrical accidents.

Authors:  Keiko Nishiyama; Makiko Komori; Mitsuharu Kodaka; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2010-08-14       Impact factor: 1.731

3.  Intrathoracic fire during preparation of the left internal thoracic artery for coronary artery bypass grafting.

Authors:  Martin Friedrich; Theodor Tirilomis; Jan D Schmitto; Aron F Popov; Suyog A Mokashi; Marc Hinterthaner; Gunnar G Hanekop; Paul Zwaka; Friedrich A Schoendube
Journal:  J Cardiothorac Surg       Date:  2010-03-10       Impact factor: 1.637

Review 4.  Electrosurgical unit: Iatrogenic injuries and medico-legal aspect. Italian legal rules, experience and article review.

Authors:  Patrizia Gualniera; Serena Scurria; Daniela Sapienza; Alessio Asmundo
Journal:  Ann Med Surg (Lond)       Date:  2021-01-04

5.  Rare case-series of electrocautery burn following off-pump coronary artery bypass grafting.

Authors:  Feridoun Sabzi; Mojtaba Niazi; Alireza Ahmadi
Journal:  J Inj Violence Res       Date:  2013-05-14
  5 in total