Oxygen delivery using neonatal self-inflating resuscitation bags without a reservoir.

M Thió1, R Bhatia, J A Dawson, P G Davis.   

Abstract

BACKGROUND: Guidelines recommend avoidance of excessive oxygen administration during neonatal resuscitation. Blenders are used in some but not all hospitals. It has been suggested that self-inflating bags without a reservoir deliver around 40% oxygen and could be used to provide an inexpensive and effective technique of avoiding oxygen toxicity.
OBJECTIVE: To explore how much oxygen is delivered when using two different brands of neonatal self-inflating resuscitation bags without a reservoir.
METHODS: In a benchtop setting, the smallest non-disposable self-inflating bags from the Laerdal and Ambu ranges were tested. Oxygen concentration delivered by these devices under a variety of conditions was measured. 108 combinations of oxygen flow rates (10; 5 to 1 litre/min), ventilation rates (30, 60, 100 inflations/min) and peak inspiratory pressure ranges (20 to 25 cm H(2)O, 35 to 40 cm H(2)O or pop-off valve range, 55 to 60 cm H(2)O) were tested.
RESULTS: Delivered oxygen concentration varied depending on three parameters: gas flow rate, ventilatory rate and pressure. At a pressure of 20 to 25 cm H(2)O, mean oxygen concentration delivered by both bags exceeded 70% at any gas flow rate except for 1 litre/min (where delivered oxygen concentration was 60% to 70%). When the pop-off valve was opened at 35 to 40 cm H(2)0, oxygen concentrations fell to 30% to 45% at gas flow rates < or =2 litres/min. The Ambu bag delivered a lower oxygen concentration than the Laerdal bag but this difference was not clinically important.
CONCLUSION: When using the Laerdal and Ambu infant resuscitation self-inflating bags without a reservoir, delivered oxygen concentration is >70% for currently recommended flow and pressure settings.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19843501     DOI: 10.1136/adc.2009.166462

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  1 in total

1.  NISHASH: A reasonable cost-effective mechanical ventilator for COVID affected patients in Bangladesh.

Authors:  Md Hafizul Imran; Rifat Bin Mahi; Rony Saha; Md Hasan Islam; Imran Mahmud
Journal:  Heliyon       Date:  2022-05-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.