| Literature DB >> 19836523 |
Juan C M Gamboa1, Roselyn C Peña, Thiago R L C Paixão, Mauro Bertotti.
Abstract
A novel approach was developed for nitrate analysis in a FIA configuration with amperometric detection (E=-0.48 V). Sensitive and reproducible current measurements were achieved by using a copper electrode activated with a controlled potential protocol. The response of the FIA amperometric method was linear over the range from 0.1 to 2.5 mmol L(-1) nitrate with a detection limit of 4.2 micromol L(-1) (S/N=3). The repeatability of measurements was determined as 4.7% (n=9) at the best conditions (flow rate: 3.0 mL min(-1), sample volume: 150 microL and nitrate concentration: 0.5 mmol L(-1)) with a sampling rate of 60 samples h(-1). The method was employed for the determination of nitrate in mineral water and soft drink samples and the results were in agreement with those obtained by using a recommended procedure. Studies towards a selective monitoring of nitrite were also performed in samples containing nitrate by carrying out measurements at a less negative potential (-0.20 V).Entities:
Mesh:
Substances:
Year: 2009 PMID: 19836523 DOI: 10.1016/j.talanta.2009.07.028
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057