The effect of serum protein concentration on wear rates in a hip simulator.

Kenneth St John1.   

Abstract

Ultra high molecular weight polyethylene cups, 22 mm in diameter, were aged for 5 years in the normal laboratory environment. Half of the samples had been processed by the standard radiation sterilization techniques, while the remainder had been cross-linked by a technique involving higher radiation doses and controlled temperature at the time of irradiation. The samples had been tested in a hip simulator for 5,000,000 cycles using a lubricant that had been diluted 1:1 with deionized water. Once that testing was completed, further testing was conducted using lubricant with greater and lesser serum protein concentrations, and the results compared with those that had already been recorded. Comparison of the wear rates within the study as well as to published data concerning the effect of serum concentration showed results that were consistent with assumed differences in lubrication ability at different concentrations. The results of other published studies were found to be inconsistent with each other and different from some of the results of this study. There is shown to be a need for carefully controlled and conducted studies to agree, if possible, on the importance of the serum concentration and the appropriate parameters to be used in testing, as well as variations that may be necessary with different bearing material characteristics.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19833674     DOI: 10.1177/0885328209347962

Source DB:  PubMed          Journal:  J Biomater Appl        ISSN: 0885-3282            Impact factor:   2.646


  2 in total

1.  How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?

Authors:  Jörn Reinders; Robert Sonntag; Jan Philippe Kretzer
Journal:  J Mater Sci Mater Med       Date:  2014-07-12       Impact factor: 3.896

2.  No adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups: an RSA study of 8 patients 10 yaers after surgery.

Authors:  Stephan M Röhrl; Bo Nivbrant; Kjell G Nilsson
Journal:  Acta Orthop       Date:  2012-01-17       Impact factor: 3.717

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.