Test-retest variability of intraocular pressure and ocular pulse amplitude for dynamic contour tonometry: a multicentre study.

P Fogagnolo1, M Figus, P Frezzotti, M Iester, F Oddone, M Zeppieri, A Ferreras, P Brusini, L Rossetti, N Orzalesi.   

Abstract

AIMS: To assess the test-retest variability of intraocular pressure (IOP) and ocular pulse amplitude (OPA) measurements utilising dynamic contour tonometry (DCT) and to evaluate possible influential factors.
METHODS: The study included 350 consecutive subjects (175 glaucoma, 175 control; one eye per subject) from seven European centres. IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by DCT (DCT1, DCT2) in a randomised sequence. OPA was also recorded for both DCT measurements. Differences (DCT1-DCT2; OPA1-OPA2; GAT-DCT1; GAT-DCT2) were assessed using the t test. The intraclass coefficient of correlation (ICC) and coefficient of variation (CoV) for DCT and OPA were calculated.
RESULTS: DCT1 was 0.6+/-1.6 mm Hg higher than DCT2 (p<0.001); OPA1 was 0.1+/-0.7 mm Hg higher than OPA2 (p=0.02). Results were not influenced by randomisation test order. In both glaucoma and normal subjects, DCT and OPA showed ICC>0.90 and >0.76, and CoV=4.8-5.0% and 10.3-10.5%, respectively. DCT1 and 2 were 2.4+/-2.6 and 1.8+/-2.6 mm Hg higher respectively than GAT (p<0.001). DISCUSSION: DCT test-retest variability was almost perfect for IOP and good for OPA. Tonometry measurements with DCT tended to be overestimated compared with GAT.

Entities:  

Mesh:

Year:  2009        PMID: 19833616     DOI: 10.1136/bjo.2009.165142

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  3 in total

1.  Repeatability and reproducibility of applanation resonance tonometry: a cross-sectional study.

Authors:  Laura Ottobelli; Paolo Fogagnolo; Paolo Frezzotti; Stefano De Cillà; Elena Vallenzasca; Maurizio Digiuni; Ruggiero Paderni; Ilaria Motolese; Simone Alex Bagaglia; Eduardo Motolese; Luca Rossetti
Journal:  BMC Ophthalmol       Date:  2015-04-10       Impact factor: 2.209

2.  Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices.

Authors:  Michele Lanza; Michele Rinaldi; Ugo Antonello Gironi Carnevale; Silvio di Staso; Mario Bifani Sconocchia; Ciro Costagliola
Journal:  BMC Ophthalmol       Date:  2018-09-03       Impact factor: 2.209

Review 3.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.