Literature DB >> 19828515

F4H5: a novel substance for the removal of silicone oil from intraocular lenses.

Theodor Stappler1, Rachel Williams, David Wong.   

Abstract

AIM: Adherent silicone oil on intraocular lenses (IOLs) following retinal detachment surgery induces large and irregular refractive errors and multiple images, and gives rise to glare, distorted and often poor vision. Its removal remains challenging, often requiring mechanical wiping or explantation. F4H5 is a new semifluorinated alkane into which silicone oil is readily soluble. The aim is to establish the effectiveness of F4H5 in removing silicone oil from three different types of IOL in vitro.
METHOD: Silicone lenses (Tecnis ZM900, Advanced Medical Optics, Inc.), hydrophobic acrylic lenses (MA60, Alcon Laboratories, Inc.) and PMMA lenses (Ocular Vision, Inc) were first immersed in phosphate-buffered saline, second in silicone oil, then in F4H5 (Fluoron GmbH) for 10 min and lastly vigorously agitated in F4H5 for 1 min. They were weighed at each stage using scales accurate to 0.0001 g to measure the weight of the adherent oil. Dynamic contact angle (DCA) analysis was used to assess their surface properties.
RESULTS: Immersion in F4H5 alone removed 96.1% (+/-1.23) by weight of silicone oil from the hydrophobic acrylic lenses, 91.4% (+/-1.58) from the silicone and 95.6% (+/-1.44) from the PMMA IOLs. Immersion combined with 1 min of agitation increased the removal to 98.8% (+/-0.46) from the acrylic IOLs, to 93.7% (+/-0.48) from the silicone IOLs and to 100% (within +/-0.0001 g) from every PMMA IOL. After treatment with F4H5, all IOL were optically clear. DCA hysteresis curves remained permanently altered. All measurements were highly reproducible.
CONCLUSION: F4H5 was highly effective at removing the bulk of the silicone oil from all three groups of IOL. The DCA measurements suggested that their surface properties were permanently modified.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19828515     DOI: 10.1136/bjo.2009.161489

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  4 in total

Review 1.  [Heavy liquids as intraoperative instrument in retinal surgery].

Authors:  Karl Boden; K Januschowski; P Szurman
Journal:  Ophthalmologe       Date:  2019-10       Impact factor: 1.059

2.  Removal of Silicone Oil From Intraocular Lens Using Novel Surgical Materials.

Authors:  Eleftherios I Paschalis; Dean Eliott; Demetrios G Vavvas
Journal:  Transl Vis Sci Technol       Date:  2014-09-12       Impact factor: 3.283

Review 3.  Biocompatibility of intraocular liquid tamponade agents: an update.

Authors:  Mario R Romano; Mariantonia Ferrara; Irene Nepita; Jana D'Amato Tothova; Alberto Giacometti Schieroni; Daniela Reami; Raniero Mendichi; Libero Liggieri; Rodolfo Repetto
Journal:  Eye (Lond)       Date:  2021-05-25       Impact factor: 4.456

4.  COHORT SAFETY AND EFFICACY STUDY OF SILURON2000 EMULSIFICATION-RESISTANT SILICONE OIL AND F4H5 IN THE TREATMENT OF FULL-THICKNESS MACULAR HOLE.

Authors:  Peter Stalmans; Anne-Marie Pinxten; David S Wong
Journal:  Retina       Date:  2015-12       Impact factor: 4.256

  4 in total