Electromagnetic interference of communication devices on ECG machines.

Adrian Baranchuk1, Jaskaran Kang, Cathy Shaw, Debra Campbell, Sebastian Ribas, Wilma M Hopman, Haitham Alanazi, Damian P Redfearn, Christopher S Simpson.   

Abstract

BACKGROUND: Use of communication devices in the hospital environment remains controversial. Electromagnetic interference (EMI) can affect different medical devices. Potential sources for EMI on ECG machines were systematically tested. HYPOTHESIS: Communication devices produce EMI on ECG machines. EMI impairs ECG interpretation.
METHODS: The communication devices tested were: a global system for mobile communication (GSM) receiver, a code division multiple access (CDMA) receiver, an analog phone, a wireless local area network, and an alpha-numeric pager. EMI was tested on 3 ECG machines: MAC 5000, MAC 1200, and ELI 100. The devices were tested at 2 and 1 meter, 50, 25, and 0 cm from the acquisition module. The ECGs were presented to a heterogeneous group of clinical providers, (medical students, residents, nurses, industry representatives from cardiac devices companies, and attending cardiologists) to evaluate the impact of EMI on ECG interpretation skills.
RESULTS: EMI was detected on the MAC 5000 ECG machine when activated GSM, CDMA, and analog phones were placed on top of the acquisition module. No EMI was seen on the other ECG machines or when phones were at a longer distance or deactivated. EMI was incorrectly diagnosed in 18% of the cases. EMI was confused most frequently with atrial fibrillation or flutter (52%), ventricular arrhythmias (22%), and pacemaker dysfunction (26%). Medical students (p < 0.003) and non-cardiology residents (p = 0.05) demonstrated significantly worse performance on EMI interpretation.
CONCLUSIONS: Digital and analog phones produce EMI on modern ECG machines when activated in direct contact to the acquisition module. EMI impairs ECG interpretation.

Entities:  

Mesh:

Year:  2009        PMID: 19824066      PMCID: PMC6652930          DOI: 10.1002/clc.20459

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   2.882


  4 in total

1.  Electromagnetic interference of wireless local area network on electrocardiogram monitoring system: a case report.

Authors:  Seungmin Chung; Joohee Yi; Seung Woo Park
Journal:  Korean Circ J       Date:  2013-03-31       Impact factor: 3.243

2.  Expert Hypertension Detection System Featuring Pulse Plethysmograph Signals and Hybrid Feature Selection and Reduction Scheme.

Authors:  Muhammad Umar Khan; Sumair Aziz; Tallha Akram; Fatima Amjad; Khushbakht Iqtidar; Yunyoung Nam; Muhammad Attique Khan
Journal:  Sensors (Basel)       Date:  2021-01-02       Impact factor: 3.576

3.  In situ diagnosis and simultaneous treatment of cardiac diseases using a single-device platform.

Authors:  Jae Chul Hwang; Moohyun Kim; Sumin Kim; Hunkyu Seo; Soohwan An; Eui Hwa Jang; Seung Yeop Han; Mi Jung Kim; Nam Kyun Kim; Seung-Woo Cho; Sak Lee; Jang-Ung Park
Journal:  Sci Adv       Date:  2022-09-14       Impact factor: 14.957

4.  Can active signals of cellphone interfere with electronic working length determination of a root canal in a dental clinic? An in vivo study.

Authors:  Umadevi Kiritsinh Gohil; Vaishali V Parekh; Niraj Kinariwala; Kaushal M Oza; Mona C Somani
Journal:  J Conserv Dent       Date:  2017 May-Jun
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.