Intrathoracic and ventricular impedances are associated with changes in ventricular volume in patients receiving defibrillators for CRT.

Massimiliano Maines1, Maurizio Landolina, Maurizio Lunati, Gabriele Lonardi, Alessia Pappone, Alessandro Proclemer, Gabriele Zanotto, Massimo Santini, Annamaria Varbaro, Marco Vimercati, Sergio Valsecchi.   

Abstract

BACKGROUND: Some implantable cardioverter defibrillators (ICD) are able to monitor intrathoracic impedance to detect pulmonary fluid overload. This is achieved by measuring impedance between the ICD case and the right ventricular (RV) lead. We hypothesized that the measured impedance would rise with improvement in left ventricular (LV) volumes during cardiac resynchronization therapy (CRT), and that such impedance changes would be more apparent when measured with an alternative pacing vector.
METHODS: We analyzed echocardiographic and impedance data from heart failure patients implanted with a CRT-ICD capable of intrathoracic impedance measurement for fluid accumulation diagnosis, and LV pacing impedance recording for lead integrity monitoring.
RESULTS: In 127 out of 170 patients that received de novo CRT implantation, the LV end-systolic volume (LVESV) decreased at 6-month follow-up (LVESV at 6 month-LVESV at baseline <0: group A). For the remaining 43 patients (group B) the change was > or = 0. Despite comparable values at baseline (P = 0.262), the impedances of groups A and B gradually diverged soon after the implant, resulting in significant difference between the two groups at the 6-month visit (P = 0.001). The changes in LV dimensions produced larger differences between groups in the impedance measured between the LV and the RV leads (P < 0.001). The regression analysis demonstrated an inverse correlation between paired changes of volume and intrathoracic impedance. Higher correlation coefficient was obtained using the LV-to-RV measurement vector (r =-0.635, P < 0.001).
CONCLUSIONS: The changes in ICD-measured impedance seem associated with the LV volume changes induced by CRT. Specifically, the LV-to-RV impedance estimations seem to better correlate with paired changes of ventricular volumes.

Entities:  

Mesh:

Year:  2009        PMID: 19821939     DOI: 10.1111/j.1540-8159.2009.02579.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  7 in total

1.  Intracardiac impedance after cardiac resynchronization therapy is a novel predictor for worsening of heart failure.

Authors:  Hitoshi Suzuki; Minoru Nodera; Masashi Kamioka; Takashi Kaneshiro; Yoshiyuki Kamiyama; Yasuchika Takeishi
Journal:  Heart Vessels       Date:  2017-02-08       Impact factor: 2.037

Review 2.  Sex differences in device therapy for heart failure: utilization, outcomes, and adverse events.

Authors:  Naomi D Herz; Joseph Engeda; Robbert Zusterzeel; William E Sanders; Kathryn M O'Callaghan; David G Strauss; Samantha B Jacobs; Kimberly A Selzman; Ileana L Piña; Daniel A Caños
Journal:  J Womens Health (Larchmt)       Date:  2015-03-20       Impact factor: 2.681

3.  Potential value of automated daily screening of cardiac resynchronization therapy defibrillator diagnostics for prediction of major cardiovascular events: results from Home-CARE (Home Monitoring in Cardiac Resynchronization Therapy) study.

Authors:  Stefan Sack; Christian Michael Wende; Herbert Nägele; Amos Katz; Wolfgang Rudolf Bauer; Craig Scott Barr; Klaus Malinowski; Harald Schwacke; Francisco Leyva; Jochen Proff; Sergey Berdyshev; Vincent Paul
Journal:  Eur J Heart Fail       Date:  2011-09       Impact factor: 15.534

4.  Estimation of Central Venous Pressure by Pacemaker Lead Impedances in Left Ventricular Assist Device Patients.

Authors:  Teruhiko Imamura; Joshua D Moss; Erin Flatley; Daniel Rodgers; Gene Kim; Jayant Raikhelkar; Nitasha Sarswat; Sara Kalantari; Ann Nguyen; Colleen Juricek; Daniel Burkhoff; Tae Song; Takeyoshi Ota; Valluvan Jeevanandam; Gabriel Sayer; Nir Uriel
Journal:  ASAIO J       Date:  2020-01       Impact factor: 3.826

5.  Thoracic impedance measures tissue characteristics in the vicinity of the electrodes, not intervening lung water: implications for heart failure monitoring.

Authors:  Christopher J Charles; Miriam T Rademaker; Iain C Melton; Dan Gutfinger; Neal L Eigler; Fujian Qu; Richard W Troughton
Journal:  J Clin Monit Comput       Date:  2014-03-12       Impact factor: 2.502

6.  Preliminary experience with a novel Multisensor algorithm for heart failure monitoring: The HeartLogic index.

Authors:  Luca Santini; Karim Mahfouz; Valentina Schirripa; Nicola Danisi; Michelangelo Leone; Gloria Mangone; Monica Campari; Sergio Valsecchi; Fabrizio Ammirati
Journal:  Clin Case Rep       Date:  2018-05-29

7.  Implantable Cardioverter Defibrillator Multisensor Monitoring during Home Confinement Caused by the COVID-19 Pandemic.

Authors:  Matteo Ziacchi; Leonardo Calò; Antonio D'Onofrio; Michele Manzo; Antonio Dello Russo; Luca Santini; Giovanna Giubilato; Cosimo Carriere; Vincenzo Ezio Santobuono; Gianluca Savarese; Carmelo La Greca; Giuseppe Arena; Antonello Talarico; Ennio Pisanò; Massimo Giammaria; Antonio Pangallo; Monica Campari; Sergio Valsecchi; Igor Diemberger
Journal:  Biology (Basel)       Date:  2022-01-12
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.