Influence of an additional locking screw on fracture reduction after volar fixed-angle plating-introduction of the "protection screw" in an extra-articular distal radius fracture model.

Patrick Weninger1, Michael Schueller, Herwig Drobetz, Michael Jamek, Heinz Redl, Elmar Tschegg.   

Abstract

BACKGROUND: Though, volar fixed-angle plating has become increasingly popular to treat distal radius fractures no studies are available comparing the mechanical properties of different screw configurations to fix the plate to the shaft. The aim of the present study was to evaluate the effect of an additional locking screw directly proximal to the fracture site and to assess if such a screw might be protective against secondary loss of reduction after volar fixed-angle plating.
METHODS: Ten Sawbones radii were used to simulate an extra-articular distal radius fracture model (AO/OTA 23-A3). In the first group (n = 5), volar fixed-angle plates (Aptus Radius 2.5, Medartis, Switzerland) were fixed to the radius shaft with a single nonlocking cortex screw in the oval hole and with two locking screws in the holes proximal to the long hole (LLNx); in the second group, the plates were fixed identically as in the first group, but an additional locking screw was inserted into the plate-hole distal to the oval hole, proximal to the fracture site (LLNL). After embedding, specimens were tested with a servohydaulic material testing machine under cyclic axial loading with 800 N for 2,000 cycles. Axial stiffness, elastic axial deformation (elastic dorsal tilt angle), and plastic deformation (plastic dorsal tilt angle) were recorded.
RESULTS: After 2,000 cycles, stiffness was 761.6 (+/-59.5) N/mm in group LLNx and 628.9 (+/-37.6) N/mm in group LLNL without significant difference. Elastic deformation and elastic tilt angle were 1.05 (+/-0.08) mm and 3.9 (+/-0.3) degree in group LLNx. In group LLNL, elastic deformation and elastic tilt angle were 1.27 (+/-0.08) mm and 4.7 (+/-0.3) degree, respectively. Plastic deformation and plastic tilt angle were significantly higher in group LLNx (p < 0.001). After 2,000 cycles, dorsal tilt angle was 1.9 (+/-0.3) degree in group LLNx and 0.7 (+/-1.0) in group LLNL.
CONCLUSION: This study showed the mechanical superiority of volar fixed-angle plates with a "protection screw" in an extra-articular fracture model. Further clinical investigation is needed to verify the results.

Entities:  

Mesh:

Year:  2009        PMID: 19820581     DOI: 10.1097/TA.0b013e3181a8b238

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  3 in total

1.  The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates.

Authors:  Lindley B Wall; Michael D Brodt; Matthew J Silva; Martin I Boyer; Ryan P Calfee
Journal:  J Hand Surg Am       Date:  2012-02-02       Impact factor: 2.230

2.  [Multidirectional locking volar plate in distal radius fractures. Secondary loss of reduction--correct implantation is essential].

Authors:  C Michelitsch; Y P Acklin; C Sommer
Journal:  Unfallchirurg       Date:  2013-12       Impact factor: 1.000

3.  The influence of distal screw length on the primary stability of volar plate osteosynthesis--a biomechanical study.

Authors:  Sebastian F Baumbach; Alexander Synek; Hannes Traxler; Wolf Mutschler; Dieter Pahr; Yan Chevalier
Journal:  J Orthop Surg Res       Date:  2015-09-08       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.