A novel device (SD-101) with high accuracy for screening sleep apnoea-hypopnoea syndrome.

Toshihiko Agatsuma1, Keisaku Fujimoto, Yoshimichi Komatsu, Kazuhisa Urushihata, Takayuki Honda, Teruomi Tsukahara, Tetsuo Nomiyama.   

Abstract

BACKGROUND AND
OBJECTIVE: The SD-101 is a non-restrictive, sheet-like medical device with an array of pressure sensors, to detect sleep-disordered breathing by sensing gravitational alterations in the body corresponding to respiratory movements. This study evaluated the accuracy of the SD-101 for screening sleep apnoea-hypopnoea syndrome (SAHS) by comparison with polysomnography.
METHODS: Nocturnal polysomnography and SD-101 monitoring were conducted simultaneously and compared in 201 patients with suspected SAHS (suspected SAHS group) and 165 male employees of a transport company (screening group).
RESULTS: Polysomnography revealed an AHI of <5, 5 < or = AHI < 15, 15 < or = AHI < 30, 30 < or = AHI < 60 and AHI > or = 60 events/h in 39, 35, 38, 68 and 21 subjects in the suspected SAHS group and 103, 34, 12, 12 and four subjects in the screening group, respectively. Central SAHS and obstructive SAHS were subsequently diagnosed in 11 (5.5%) and 135 (67.2%) of subjects in the suspected SAHS group and five (3.0%) and 39 (23.6%) of subjects in the screening group, respectively. Significant correlations were apparent between AHI and the respiratory disturbance index (RDI) measured with the SD-101 in both the suspected SAHS group (r = 0.88) and screening group (r = 0.92). Receiver operating characteristic curve analysis revealed 89.5% sensitivity and 85.8% specificity in identifying SAHS, using an RDI of 14.0 events/h.
CONCLUSIONS: These findings suggest that the SD-101 is a useful device for screening SAHS.

Entities:  

Mesh:

Year:  2009        PMID: 19818056     DOI: 10.1111/j.1440-1843.2009.01627.x

Source DB:  PubMed          Journal:  Respirology        ISSN: 1323-7799            Impact factor:   6.424


  7 in total

1.  Validation of the Sonomat: a contactless monitoring system used for the diagnosis of sleep disordered breathing.

Authors:  Mark B Norman; Sally Middleton; Odette Erskine; Peter G Middleton; John R Wheatley; Colin E Sullivan
Journal:  Sleep       Date:  2014-09-01       Impact factor: 5.849

2.  Accuracy of ECG-based screening for sleep-disordered breathing: a survey of all male workers in a transport company.

Authors:  Junichiro Hayano; Teruomi Tsukahara; Eiichi Watanabe; Fumihiko Sasaki; Kiyohiro Kawai; Hiroki Sakakibara; Itsuo Kodama; Tetsuo Nomiyama; Keisaku Fujimoto
Journal:  Sleep Breath       Date:  2012-03-20       Impact factor: 2.816

3.  Validity of sheet-type portable monitoring device for screening obstructive sleep apnea syndrome.

Authors:  Mina Kobayashi; Kazuyoshi Namba; Satoru Tsuiki; Masaki Nakamura; Masamichi Hayashi; Yuuki Mieno; Hiromi Imizu; Shiho Fujita; Atsushi Yoshikawa; Hiroki Sakakibara; Yuichi Inoue
Journal:  Sleep Breath       Date:  2012-06-16       Impact factor: 2.816

4.  Digital Health and Sleep-Disordered Breathing: A Systematic Review and Meta-Analysis.

Authors:  Talita Rosa; Kersti Bellardi; Alonço Viana; Yifei Ma; Robson Capasso
Journal:  J Clin Sleep Med       Date:  2018-09-15       Impact factor: 4.062

5.  Screening for obstructive sleep apnea using a contact-free system compared with polysomnography.

Authors:  Rui Zhao; JianBo Xue; Xiao Song Dong; Hui Zhi; JiaNan Chen; Long Zhao; XueLi Zhang; Jing Li; Thomas Penzel; Fang Han
Journal:  J Clin Sleep Med       Date:  2021-05-01       Impact factor: 4.062

6.  Validation of a novel sleep-monitoring system for diagnosing obstructive sleep apnea: A comparison with polysomnography.

Authors:  Lili Meng; Huajun Xu; Jian Guan; Hongliang Yi; Hongmin Wu; Shankai Yin
Journal:  Exp Ther Med       Date:  2016-09-20       Impact factor: 2.447

7.  Instability of nocturnal parasympathetic nerve function in patients with chronic lung disease with or without nocturnal desaturation.

Authors:  Keisaku Fujimoto; Haruna Yamazaki; Akikazu Uematsu
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-09-11
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.