| Literature DB >> 19817539 |
Yasunari Otawara1, Kuniaki Ogasawara, Hiroshi Kashimura, Yoshitaka Kubo, Akira Ogawa, Kouichi Watanabe.
Abstract
The mechanical properties of titanium-alloy aneurysm clips after long-term implantation in the human cranium are unclear. The characteristics of a Yasargil titanium aneurysm clip were evaluated after long-term implantation for 12 years in a patient with a cerebral aneurysm. The closing forces of the retrieved clip before and after implantation were approximately equal. The bending test showed no differences between the retrieved and control clips. Titanium oxide and calcium were identified on the surface of the retrieved clip, which indicated the formation of corrosion-resistant layers. Titanium-alloy clips retain their mechanical properties in the human cranium for a long time.Entities:
Mesh:
Substances:
Year: 2010 PMID: 19817539 DOI: 10.3171/2009.9.JNS09827
Source DB: PubMed Journal: J Neurosurg ISSN: 0022-3085 Impact factor: 5.115