Literature DB >> 19817539

Mechanical characteristics and surface elemental composition of a Yasargil titanium aneurysm clip after long-term implantation.

Yasunari Otawara1, Kuniaki Ogasawara, Hiroshi Kashimura, Yoshitaka Kubo, Akira Ogawa, Kouichi Watanabe.   

Abstract

The mechanical properties of titanium-alloy aneurysm clips after long-term implantation in the human cranium are unclear. The characteristics of a Yasargil titanium aneurysm clip were evaluated after long-term implantation for 12 years in a patient with a cerebral aneurysm. The closing forces of the retrieved clip before and after implantation were approximately equal. The bending test showed no differences between the retrieved and control clips. Titanium oxide and calcium were identified on the surface of the retrieved clip, which indicated the formation of corrosion-resistant layers. Titanium-alloy clips retain their mechanical properties in the human cranium for a long time.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19817539     DOI: 10.3171/2009.9.JNS09827

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  2 in total

1.  The Comparative Study of Microsurgical Cerebral Aneurysm Clip Implants; Titanium Clip vs. Stainless Steel Clip.

Authors:  Sung Ho Choi; Cheol Wan Park; Young Bo Kim; Eun Young Kim; Chang Jong You; Woo Kyung Kim
Journal:  J Cerebrovasc Endovasc Neurosurg       Date:  2012-06-30

2.  Multiple visceral artery aneurysms managed by Yasargil aneurysm clips.

Authors:  In-Kee Hong; Ji-Ho Choi; Young Chae Chu; Yong Sun Jeon; Keon-Young Lee
Journal:  Ann Surg Treat Res       Date:  2015-08-24       Impact factor: 1.859

  2 in total