André van Elsäcker1, Tanja Nikolic, Mike G Scheffer, Berry M van Gelder.
Abstract
Programming maximum right ventricular output in a patient with a biventricular implanted cardioverter defibrillator resulted in ventricular oversensing and ventricular safety pacing in the same cardiac cycle.Entities:
Mesh:
Year: 2010 PMID: 19801565 DOI: 10.1093/europace/eup283