Cyclic testing of 3 all-inside meniscal repair devices: a biomechanical analysis.

Vishal M Mehta1, Michael A Terry.   

Abstract

BACKGROUND: Despite the growing popularity of all-inside meniscal repair devices, concerns remain about their fixation strength. It is also unclear which of these devices have the most ideal biomechanical properties.
PURPOSE: To compare the biomechanical properties of 3 all-inside meniscal repair devices: the Meniscal Cinch, Ultra FAST-FIX, and MaxFire. STUDY
DESIGN: Controlled laboratory study.
METHODS: Twenty-seven human cadaveric menisci (3 groups of 9) were repaired using 3 different meniscal repair devices. The repaired menisci were then subjected to cyclic loading and load-to-failure testing. Gap formation and ultimate load to failure were measured.
RESULTS: Six of the devices failed during cyclic testing, 4 in the MaxFire group (44%), 1 in the Ultra FAST-FIX group (11%), and 1 in the Meniscal Cinch group (11%). After 1 cycle, there was a trend toward larger gap formation in the MaxFire group (3.65 mm) compared with the Meniscal Cinch group (2.12 mm, P = .05). After 100 cycles, group differences were found in gap formation (P = .03), with the MaxFire group exhibiting greater displacement (6.70 mm) than the Ultra FAST-FIX group (3.59 mm). After 500 cycles, group differences in gap formation (Meniscal Cinch, 5.94; Ultra FAST-FIX, 4.74 mm; Max Fire, 7.19 mm) did not reach statistical significance (P = .20). A trend was found toward higher ultimate load to failure in the Ultra FAST-FIX (86.1 N) and Meniscal Cinch (85.3 N) groups compared with the MaxFire group (64.5 N, P = .06). Stiffness was also higher in the Ultra FAST-FIX (25.2 N/mm) and Meniscal Cinch (25.5 N/mm) groups than the MaxFire group (16.3 N/mm, P = .02).
CONCLUSION: The Meniscal Cinch and Ultra FAST-FIX devices have more desirable biomechanical properties than the MaxFire as demonstrated by higher stiffness and a lower failure rate during cyclic testing. CLINICAL RELEVANCE: The Meniscal Cinch and Ultra FAST-FIX devices may be more desirable implants for use during all-inside meniscal repair as they have superior biomechanical properties when compared with the MaxFire device.

Entities:  

Mesh:

Year:  2009        PMID: 19797615     DOI: 10.1177/0363546509346051

Source DB:  PubMed          Journal:  Am J Sports Med        ISSN: 0363-5465            Impact factor:   6.202


  8 in total

1.  Mechanical comparison of meniscal repair devices with mattress suture devices in vitro.

Authors:  Brian C Aros; Angela Pedroza; William K Vasileff; Alan S Litsky; David C Flanigan
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-06-09       Impact factor: 4.342

2.  Arthroscopically Repaired Bucket-Handle Meniscus Tears: Patient Demographics, Postoperative Outcomes, and a Comparison of Success and Failure Cases.

Authors:  Bryan M Saltzman; Eric J Cotter; Kevin C Wang; Richard Rice; Blaine T Manning; Adam B Yanke; Brian Forsythe; Nikhil N Verma; Brian J Cole
Journal:  Cartilage       Date:  2018-06-29       Impact factor: 4.634

3.  The concept of three-dimensional hold of both circumferential and radial collagen fibres of the meniscus.

Authors:  Ashraf Abdelkafy
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-02-12       Impact factor: 4.342

Review 4.  Meta-analysis on biomechanical properties of meniscus repairs: are devices better than sutures?

Authors:  Daniel M Buckland; Patrick Sadoghi; Matthias D Wimmer; Patrick Vavken; Geert I Pagenstert; Victor Valderrabano; Claudio Rosso
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-05-20       Impact factor: 4.342

5.  Bonding and fusion of meniscus fibrocartilage using a novel chondroitin sulfate bone marrow tissue adhesive.

Authors:  Jacob A Simson; Iossif A Strehin; Brian W Allen; Jennifer H Elisseeff
Journal:  Tissue Eng Part A       Date:  2013-06-12       Impact factor: 3.845

6.  Biomechanical Comparison of Parallel and Crossed Suture Repair for Longitudinal Meniscus Tears.

Authors:  Charles Milchteim; Eric A Branch; Ty Maughon; Jay Hughey; Adam W Anz
Journal:  Orthop J Sports Med       Date:  2016-04-08

7.  Evaluation of Midterm Clinical and Radiographic Outcomes of Arthroscopically Repaired Vertical Longitudinal and Bucket-Handle Lateral Meniscal Tears.

Authors:  Erdal Uzun; Abdulhamit Misir; Turan Bilge Kizkapan; Mustafa Ozcamdalli; Soner Akkurt; Ahmet Guney
Journal:  Orthop J Sports Med       Date:  2019-05-10

8.  Increased Construct Stiffness With Meniscal Repair Sutures and Devices Increases the Risk of Cheese-Wiring During Biomechanical Load-to-Failure Testing.

Authors:  Sebastian Müller; Tanja Schwenk; Michael de Wild; Dimitris Dimitriou; Claudio Rosso
Journal:  Orthop J Sports Med       Date:  2021-06-15
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.