Omkar Karhade1, Levent Degertekin, Thomas Kurfess.
Abstract
An active control method is proposed and demonstrated to improve grating-based laser interferometry to achieve low-noise (subpicometer resolution) and multiwavelength unambiguous range of operation simultaneously. The method modifies a recurrent calibration-based path stabilization algorithm to extract high-resolution and low-resolution data in parallel. This extended range recurrent calibration method is experimentally verified by implementing it on the micromachined scanning grating interferometer (microSGI).Year: 2009 PMID: 19794810 DOI: 10.1364/OL.34.003044
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776