Literature DB >> 19793371

Primary prevention of atrial fibrillation: does the atrial lead position influence the incidence of atrial arrhythmias in patients with sinus node dysfunction? Results from the PASTA Trial.

Stefan G Spitzer1, Petra Wacker, Steffen Gazarek, Klaus Malinowski, Volker Schibgilla.   

Abstract

INTRODUCTION: PASTA (pacing of the atria in sinus node disease) is a prospective and randomized trial, assessing the effect of different atrial lead positions on the atrial fibrillation (AF) incidence in patients with sinus node disease (SND).
METHOD: The atrial lead position is randomized to: (a) free right atrial wall, (b) right atrial appendage (RAA), (c) coronary sinus ostium (CS-Os), or (d) dual site right atrial pacing (CS-Os + RAA). The pacemakers (Vitatron Selection 9000 or Prevent AF, Vitatron B.V., Arnhem, The Netherlands) are programmed in DDDR 70 mode and the total follow-up duration is 24 months. To describe the atrial rhythm state, pacemaker-derived data (arrhythmia counter) were assessed for AF episodes. AF was considered as evident, if the AF burden (time in AF related to follow-up interval) was >1% (i.e., 15 min/d). Follow-up data after 24 months were evaluated.
RESULTS: The analysis evaluates 142 patients (77 male, 74.5 +/- 7.8 years). There was no statistical significant difference with respect to the occurrence of AF between the four groups after 24 months (A: 36%; B: 38%, C: 32%, D: 48%). The percentage of atrial/ventricular pacing was in A: 78/76%, in B: 84/81%, in C: 70/65%, and in D: 79/69%. These differences were not significant.
CONCLUSION: The evaluation of the AF burden >1% and the total AF burden after 24 months did not show differences in the incidence of AF in patients with dual chamber pacemaker therapy for SND. We were not able to demonstrate a significant influence of right atrial lead position on the incidence of AF recurrence.

Entities:  

Mesh:

Year:  2009        PMID: 19793371     DOI: 10.1111/j.1540-8159.2009.02544.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Interatrial septum versus right atrial appendage pacing for prevention of atrial fibrillation : A meta-analysis of randomized controlled trials.

Authors:  L Zhang; H Jiang; W Wang; J Bai; Y Liang; Y Su; J Ge
Journal:  Herz       Date:  2017-07-28       Impact factor: 1.443

Review 2.  [Bachmann bundle pacing].

Authors:  C W Israel; T Burmistrava; C Berger
Journal:  Herz       Date:  2018-11       Impact factor: 1.443

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.