Vacuum-compatible pulse selector for free-electron laser.

Togo Kudo1, Toko Hirono, Mitsuru Nagasono, Makina Yabashi.   

Abstract

We developed a vacuum-compatible pulse selector for a free-electron laser. A rotating cylinder with eight apertures in a vacuum is driven by a closed-loop stepping motor system through a magnetically coupled rotary feedthrough. A field programmable gate array is used to synchronize the cylinder rotation with a trigger signal of the accelerator at a maximum repetition rate of 60 Hz. We achieved to select specific pulses from a continuous pulse train of the SPring-8 Compact SASE Source.

Year:  2009        PMID: 19791933     DOI: 10.1063/1.3223972

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  3 in total

1.  Multiple-beamline operation of SACLA.

Authors:  Kensuke Tono; Toru Hara; Makina Yabashi; Hitoshi Tanaka
Journal:  J Synchrotron Radiat       Date:  2019-02-22       Impact factor: 2.616

2.  A new concept for temporal gating of synchrotron X-ray pulses.

Authors:  D Schmidt; R Bauer; S Chung; D Novikov; M Sander; J E Pudell; M Herzog; D Pfuetzenreuter; J Schwarzkopf; R Chernikov; P Gaal
Journal:  J Synchrotron Radiat       Date:  2021-02-05       Impact factor: 2.616

3.  Development of a single-shot CCD-based data acquisition system for time-resolved X-ray photoelectron spectroscopy at an X-ray free-electron laser facility.

Authors:  Masaki Oura; Tatsuya Wagai; Ashish Chainani; Jun Miyawaki; Hiromi Sato; Masaharu Matsunami; Ritsuko Eguchi; Takayuki Kiss; Takashi Yamaguchi; Yasuhiro Nakatani; Tadashi Togashi; Tetsuo Katayama; Kanade Ogawa; Makina Yabashi; Yoshihito Tanaka; Yoshiki Kohmura; Kenji Tamasaku; Shik Shin; Tetsuya Ishikawa
Journal:  J Synchrotron Radiat       Date:  2013-12-10       Impact factor: 2.616

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.