A novel and sensitive method for ethinylestradiol quantification in human plasma by high-performance liquid chromatography coupled to atmospheric pressure photoionization (APPI) tandem mass spectrometry: application to a comparative pharmacokinetics study.

Ney Carter Borges1, Rafael Barrientos Astigarraga, Carlos Eduardo Sverdloff, Paulo Rabelo Galvinas, Washington Moreira da Silva, Vinícius Marcondes Rezende, Ronilson Agnaldo Moreno.   

Abstract

In the present study, a novel, fast, sensitive and robust method to quantify ethinylestradiol in human plasma using 17alpha-ethinylestradiol-d4 as the internal standard (IS) is described. The analyte and the IS were extracted from acidified plasma by liquid-liquid extraction (LLE) using diethyl ether-hexane followed by online solid phase extraction (SPE) using online C18 cartridges. Extracted samples were analyzed by high-performance liquid chromatography coupled to atmospheric pressure photoionization tandem mass spectrometry (HPLC-APPI-MS/MS). Chromatography was performed isocratically on a C18, 5 microm analytical column. The method had a chromatographic run time of 2.50 min and a linear calibration curve over the range 5-500 pg ml(-1) (r(2)>0.9992). The lowest concentration quantified was 5 pg ml(-1), demonstrating acceptable accuracy and precision. The intra-assay precisions ranged from 2.1 to 14.6%, while inter-assay precisions ranged from 4.4 to 11.4%. The intra-assay accuracies ranged from 94.6 to 103.8%, while the inter-assay accuracies ranged from 98.9 to 101.6%. The recovery of ethinylestradiol was determined as part of the assay validation process and was 73.1 and 79.0% for the concentrations 15 and 375 pg ml(-1), respectively. Short-term stability showed that ethinylestradiol was stable in plasma for at least 19 h at room temperature or for at least 385 days when stored at -20 degrees C. In the study of bioequivalence conducted in Brazil, healthy volunteers received two ethinylestradiol 0.035 mg tablet formulations using an open, randomized, two-period crossover design with a 2-week washout interval. Since the 90% confidence interval for C(max) and area under the curve ratios were all inside the 80-125% interval proposed by the US Food and Drug Administration, it was concluded that the two ethinylestradiol formulations are bioequivalent with respect to both the rate and the extent of absorption.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19773204     DOI: 10.1016/j.jchromb.2009.08.048

Source DB:  PubMed          Journal:  J Chromatogr B Analyt Technol Biomed Life Sci        ISSN: 1570-0232            Impact factor:   3.205


  4 in total

1.  In vivo D2O labeling to quantify static and dynamic changes in cholesterol and cholesterol esters by high resolution LC/MS.

Authors:  Jose Castro-Perez; Stephen F Previs; David G McLaren; Vinit Shah; Kithsiri Herath; Gowri Bhat; Douglas G Johns; Sheng-Ping Wang; Lyndon Mitnaul; Kristian Jensen; Robert Vreeken; Thomas Hankemeier; Thomas P Roddy; Brian K Hubbard
Journal:  J Lipid Res       Date:  2010-09-30       Impact factor: 5.922

2.  Liver involvement in hereditary hemorrhagic telangiectasia: can breath test unmask impaired hepatic first-pass effect?

Authors:  Marcello Candelli; Maurizio Pompili; Patrizia Suppressa; Gennaro M Lenato; Giulia Bosco; Gian Ludovico Rapaccini; Antonio Gasbarrini; Arnaldo Scardapane; Carlo Sabbà
Journal:  Intern Emerg Med       Date:  2011-02-09       Impact factor: 3.397

3.  Incorporation of concentration data below the limit of quantification in population pharmacokinetic analyses.

Authors:  Ron J Keizer; Robert S Jansen; Hilde Rosing; Bas Thijssen; Jos H Beijnen; Jan H M Schellens; Alwin D R Huitema
Journal:  Pharmacol Res Perspect       Date:  2015-03-25

4.  High-sensitivity simultaneous liquid chromatography-tandem mass spectrometry assay of ethinyl estradiol and levonorgestrel in human plasma.

Authors:  Abhishek Gandhi; Swati Guttikar; Priti Trivedi
Journal:  J Pharm Anal       Date:  2015-02-20
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.