| Literature DB >> 19771597 |
Johan W van Neck1, B Stefan de Kool, J Ineke M Hekking-Weijma, Erik T Walbeehm, Gerhard H Visser, Joleen H Blok.
Abstract
Ultrasound-guided near-nerve neurography is a new tool that can be used to assess nerve regeneration before reinnervation occurs. In this study, ultrasound-guided near-nerve measurements were validated against axon diameter counts in rabbits during a 15-week regeneration period after a crush lesion of their peroneal nerve. The course of the nerve was determined ultrasonically, and the active near-nerve needle electrode was maneuvered just next to the nerve under ultrasound guidance. Measured action potentials were compared with axon diameter counts from histological sections of these same nerves. A moderate to good positive correlation was found, which reached a maximum of 0.7 at a cut-off of 3 microm, corresponding to the minimal size of the myelinated axons. Our results suggest that, following a similar validation study in humans, ultrasound-guided near-nerve neurography may be clinically useful when early evaluation of nerve activity is needed.Entities:
Mesh:
Year: 2009 PMID: 19771597 DOI: 10.1002/mus.21405
Source DB: PubMed Journal: Muscle Nerve ISSN: 0148-639X Impact factor: 3.217