Modifying peripheral IV catheters with side holes and side slits results in favorable changes in fluid dynamic properties during the injection of iodinated contrast material.

Paul W Weber1, Courtney A Coursey, Laurens E Howle, Rendon C Nelson, Eli B Nichols, Sebastian T Schindera.   

Abstract

OBJECTIVE: The purpose of this study was to compare a standard peripheral end-hole angiocatheter with those modified with side holes or side slits using experimental optical techniques to qualitatively compare the contrast material exit jets and using numeric techniques to provide flow visualization and quantitative comparisons.
MATERIALS AND METHODS: A Schlieren imaging system was used to visualize the angiocatheter exit jet fluid dynamics at two different flow rates. Catheters were modified by drilling through-and-through side holes or by cutting slits into the catheters. A commercial computational fluid dynamics package was used to calculate numeric results for various vessel diameters and catheter orientations.
RESULTS: Experimental images showed that modifying standard peripheral IV angiocatheters with side holes or side slits qualitatively changed the overall flow field and caused the exiting jet to become less well defined. Numeric calculations showed that the addition of side holes or slits resulted in a 9-30% reduction of the velocity of contrast material exiting the end hole of the angiocatheter. With the catheter tip directed obliquely to the wall, the maximum wall shear stress was always highest for the unmodified catheter and was always lowest for the four-side-slit catheter.
CONCLUSION: Modified angiocatheters may have the potential to reduce extravasation events in patients by reducing vessel wall shear stress.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19770318     DOI: 10.2214/AJR.09.2521

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  4 in total

1.  Computational analysis of the effectiveness of blood flushing with saline injection from an intravascular diagnostic catheter.

Authors:  Narugopal Ghata; Ralph C Aldredge; Julien Bec; Laura Marcu
Journal:  Int J Numer Method Biomed Eng       Date:  2014-07-17       Impact factor: 2.747

2.  Acute compartment syndrome due to extravasation of peripheral intravenous blood transfusion.

Authors:  Chanyang Park; Hyuckgoo Kim
Journal:  Saudi J Anaesth       Date:  2020-03-05

3.  Impending Upper Arm Compartment Syndrome Secondary to Intravenous Fluid Infiltration.

Authors:  Amr Tawfik; Bryan Hozack; Justin Melendez; Bobby Varghese; Brian M Katt; Pedro Beredjiklian; Michael Nakashian
Journal:  Cureus       Date:  2021-06-15

4.  More holes, more contrast? Comparing an 18-gauge non-fenestrated catheter with a 22-gauge fenestrated catheter for cardiac CT.

Authors:  Andreas Marco Fischer; Philipp Riffel; Thomas Henzler; U Joseph Schoepf; Andres F Abadia; Richard Robert Bayer; Holger Haubenreisser; Dante Giovagnoli; Alexander Kremer; Stefan O Schoenberg; Joshua Gawlitza
Journal:  PLoS One       Date:  2020-06-08       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.