Evaristo Castedo1, Santiago Serrano-Fiz, Juan F Oteo, Sebastian Ramis, Paloma Martinez, Juan Ugarte.
Abstract
We report the progression of aortic insufficiency after percutaneous closure of an aortic prosthesis paravalvular leak with the Amplatzer vascular plug (AGA Inc, Golden Valley, MN). Removal of the device and replacement of the aortic prosthesis was successfully performed. Based on operative findings, we hypothesize that shape mismatch between the occluder system and the leak might promote tearing at the end of slanted defects further enhancing the regurgitant area.Entities:
Mesh:
Year: 2009 PMID: 19766832 DOI: 10.1016/j.athoracsur.2009.02.054