Literature DB >> 19760835

Failed air supply to individually ventilated caging system causes acute hypoxia and mortality of rats.

Michael J Huerkamp1, Wesley D Thompson, Noel D M Lehner.   

Abstract

Greater than 30% mortality was observed unexpectedly in rat pups of weaning age kept with their dams in individually ventilated cages (IVC). A subsequent controlled study involving an induced electrical power failure that halted ventilation of IVC containing rat dams and pups showed that oxygen concentrations in the air in individual cages were rapidly depleted to less than 10% in 1 h. In a separate incident during which electrical power to IVC was inadvertently disconnected, rats died within 60 min of loss of power to IVC. The loss of electrical power to IVC housing rats is potentially a life-threatening condition requiring an immediate response and affirms the consideration of emergency (back-up) electrical power where these systems are used.

Entities:  

Mesh:

Year:  2003        PMID: 19760835

Source DB:  PubMed          Journal:  Contemp Top Lab Anim Sci        ISSN: 1060-0558


  4 in total

1.  Growing male rats in individually ventilated and open-top cages.

Authors:  Nikolaos Kostomitsopoulos; Ismene A Dontas; Pavlos Alexakos; Pavlos Lelovas; Antonios Galanos; Euthimios Paronis; Evangelos Balafas; Konstantinos Paschidis; Alkiviadis Kostakis
Journal:  J Am Assoc Lab Anim Sci       Date:  2011-11       Impact factor: 1.232

2.  Carbon dioxide and oxygen levels in disposable individually ventilated cages after removal from mechanical ventilation.

Authors:  Claude M Nagamine; C Tyler Long; Gabriel P McKeon; Stephen A Felt
Journal:  J Am Assoc Lab Anim Sci       Date:  2012-03       Impact factor: 1.232

3.  Nasal Histopathology and Intracage Ammonia Levels in Female Groups and Breeding Mice Housed in Static Isolation Cages.

Authors:  Angela M Mexas; Angela K Brice; Adam C Caro; Troy S Hillanbrand; Diane J Gaertner
Journal:  J Am Assoc Lab Anim Sci       Date:  2015-09       Impact factor: 1.232

Review 4.  Microbiological monitoring in individually ventilated cage systems.

Authors:  Susan R Compton; Felix R Homberger; Judy MacArthur Clark
Journal:  Lab Anim (NY)       Date:  2004-11       Impact factor: 12.625

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.