Measurements of attenuation coefficient for evaluating the hardness of a cataract lens by a high-frequency ultrasonic needle transducer.

Chih-Chung Huang1, Ruimin Chen, Po-Hsiang Tsui, Qifa Zhou, Mark S Humayun, K Kirk Shung.   

Abstract

A cataract is a clouding of the lens in the eye that affects vision. Phacoemulsification is the mostly common surgical method for treating cataracts, and determining that the optimal phacoemulsification energy is dependent on measuring the hardness of the lens. This study explored the use of an ultrasound needle transducer for invasive measurements of ultrasound attenuation coefficient to evaluate the hardness of the cataract lens. A 47 MHz high-frequency needle transducer with a diameter of 0.9 mm was fabricated by a polarized PMN-33%PT single crystal in the present study. The attenuation coefficients at different stages of an artificial porcine cataract lens were measured using the spectral shift approach. The hardness of the cataract lens was also evaluated by mechanical measurement of its elastic properties. The results demonstrated that the ultrasonic attenuation coefficient was increased from 0.048 +/- 0.02 to 0.520 +/- 0.06 dB mm(-1) MHz(-1) corresponding to an increase in Young's modulus from 6 +/- 0.4 to 96 +/- 6.2 kPa as the cataract further developed. In order to evaluate the feasibility of combining needle transducer and phacoemulsification probe for real-time measurement during cataract surgery, the needle transducer was mounted on the phacoemulsification probe for a vibration test. The results indicated that there was no apparent damage to the tip of the needle transducer and the pulse-echo test showed that a good performance in sensitivity was maintained after the vibration test.

Entities:  

Mesh:

Year:  2009        PMID: 19759408      PMCID: PMC2843554          DOI: 10.1088/0031-9155/54/19/021

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  12 in total

1.  Study of the ultrasonic characteristics of the lens.

Authors:  Y Oguchi; G W van Marle; F Eijskoot; H E Henkes
Journal:  Bibl Ophthalmol       Date:  1975

2.  Determination of the velocity of ultrasound in the human lens and vitreous.

Authors:  F JANSSON; E KOCK
Journal:  Acta Ophthalmol (Copenh)       Date:  1962

3.  Absorption of ultrasound in the aqueous humour, lens and vitreous body.

Authors:  A OKSALA; A LEHTINEN
Journal:  Acta Ophthalmol (Copenh)       Date:  1958

4.  Determining the acoustic properties of the lens using a high-frequency ultrasonic needle transducer.

Authors:  Chih-Chung Huang; Qifa Zhou; Hossein Ameri; Da Wei Wu; Lei Sun; Shyh-Hau Wang; Mark S Humayun; K Kirk Shung
Journal:  Ultrasound Med Biol       Date:  2007-07-30       Impact factor: 2.998

5.  Attenuation estimation using spectral cross-correlation.

Authors:  Hyungsuk Kim; Tomy Varghese
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2007-03       Impact factor: 2.725

6.  Ultrasonic attenuation and absorption in liver tissue.

Authors:  K J Parker
Journal:  Ultrasound Med Biol       Date:  1983 Jul-Aug       Impact factor: 2.998

7.  The history of ultrasound techniques in ophthalmology.

Authors:  J M Thijssen
Journal:  Ultrasound Med Biol       Date:  1993       Impact factor: 2.998

8.  Acoustic velocity and attenuation of eye tissues at 20 MHz.

Authors:  C L de Korte; A F van der Steen; J M Thijssen
Journal:  Ultrasound Med Biol       Date:  1994       Impact factor: 2.998

9.  Hardness and ultrasonic characteristics of the human crystalline lens.

Authors:  H Tabandeh; M Wilkins; G Thompson; D Nassiri; A Karim
Journal:  J Cataract Refract Surg       Date:  2000-06       Impact factor: 3.351

10.  Evaluation of lens hardness in cataract surgery using high-frequency ultrasonic parameters in vitro.

Authors:  Chih-Chung Huang; Hossein Ameri; Charles Deboer; Adrian P Rowley; Xiaochen Xu; Lei Sun; Shyh-Hau Wang; Mark S Humayun; K Kirk Shung
Journal:  Ultrasound Med Biol       Date:  2007-07-06       Impact factor: 2.998

View more
  10 in total

1.  Thermal safety of ultrasound-enhanced ocular drug delivery: A modeling study.

Authors:  Marjan Nabili; Craig Geist; Vesna Zderic
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  Preoperative evaluation of human crystalline lens hardness using A-scan ultrasound biometry: a pilot study.

Authors:  Konstantinos T Tsaousis; Lampros P Lamprogiannis; Stavros A Dimitrakos; Ioannis T Tsinopoulos
Journal:  Int J Ophthalmol       Date:  2016-10-18       Impact factor: 1.779

3.  Relaxor-based ferroelectric single crystals: growth, domain engineering, characterization and applications.

Authors:  Enwei Sun; Wenwu Cao
Journal:  Prog Mater Sci       Date:  2014-08-01

4.  Cataract measurement by estimating the ultrasonic statistical parameter using an ultrasound needle transducer: an in vitro study.

Authors:  Po-Hsiang Tsui; Chih-Chung Huang; Qifa Zhou; K Kirk Shung
Journal:  Physiol Meas       Date:  2011-03-21       Impact factor: 2.833

5.  High-resolution acoustic-radiation-force-impulse imaging for assessing corneal sclerosis.

Authors:  Cho-Chiang Shih; Chih-Chung Huang; Qifa Zhou; K Kirk Shung
Journal:  IEEE Trans Med Imaging       Date:  2013-04-08       Impact factor: 10.048

6.  Entropic imaging of cataract lens: an in vitro study.

Authors:  Zhuhuang Zhou; Chih-Chung Huang; K Kirk Shung; Po-Hsiang Tsui; Jui Fang; Hsiang-Yang Ma; Shuicai Wu; Chung-Chih Lin
Journal:  PLoS One       Date:  2014-04-23       Impact factor: 3.240

7.  Implementation of a rotational ultrasound biomicroscopy system equipped with a high-frequency angled needle transducer--ex vivo ultrasound imaging of porcine ocular posterior tissues.

Authors:  Tae-Hoon Bok; Juho Kim; Jinho Bae; Chong Hyun Lee; Dong-Guk Paeng
Journal:  Sensors (Basel)       Date:  2014-09-24       Impact factor: 3.576

8.  Fabrication of a PMN-PT single crystal-based transcranial Doppler transducer and the power regulation of its detection system.

Authors:  Qingwen Yue; Dongxu Liu; Wei Wang; Wenning Di; Di Lin; Xi'an Wang; Haosu Luo
Journal:  Sensors (Basel)       Date:  2014-12-19       Impact factor: 3.576

9.  High-Resolution Shear Wave Imaging of the Human Cornea Using a Dual-Element Transducer.

Authors:  Pei-Yu Chen; Cho-Chiang Shih; Wei-Chen Lin; Teng Ma; Qifa Zhou; K Kirk Shung; Chih-Chung Huang
Journal:  Sensors (Basel)       Date:  2018-12-03       Impact factor: 3.576

10.  Feasibility assessment of the Eye Scan Ultrasound System for cataract characterization and optimal phacoemulsification energy estimation: protocol for a pilot, nonblinded and monocentre study.

Authors:  Lorena Petrella; Sandrina Nunes; Fernando Perdigão; Marco Gomes; Mário Santos; Carlos Pinto; Miguel Morgado; António Travassos; Jaime Santos; Miguel Caixinha
Journal:  Pilot Feasibility Stud       Date:  2022-09-29
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.