Placement of the nebulizer before the humidifier during mechanical ventilation: Effect on aerosol delivery.

Jean-Jacques Moraine1, Karine Truflandier, Nicolas Vandenbergen, Jacques Berré, Christian Mélot, Jean-Louis Vincent.   

Abstract

BACKGROUND: Therapeutic aerosols are commonly used in mechanically ventilated patients. The position of the nebulizer in the ventilator circuit and the humidification of inhaled gases can influence the efficiency of aerosol delivery. We evaluated the effect of nebulizer position on the pulmonary bioavailability of nebulized ipratropium in ventilated patients without known preexisting respiratory disease.
METHODS: The study included 38 mechanically ventilated and sedated patients after open heart surgery. Ipratropium (500 microg) was delivered by an ultrasonic nebulizer. Patients were randomized into 2 groups: the nebulizer positioned before the heat humidification system (group 1, n = 19) or at the end of the inspiratory limb before the Y-piece (group 2, n = 19). The amount of ipratropium in the urine collected during the 4 hours after drug administration was measured by mass spectrometry.
RESULTS: There were no statistically significant differences in tidal volume or respiratory rate between groups. There were no significant differences between the 2 groups in the amount of drug excreted (group 1 vs 2: 13,237 +/- 2313 pg/mL vs 15,529 +/- 3204 pg/mL) or in pulmonary bioavailability (.9% +/- .1% vs 1.1% +/- .2%).
CONCLUSION: The position of the nebulizer in the ventilatory circuit had no effect on the pulmonary bioavailability of ipratropium.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19755194     DOI: 10.1016/j.hrtlng.2008.12.005

Source DB:  PubMed          Journal:  Heart Lung        ISSN: 0147-9563            Impact factor:   2.210


  4 in total

Review 1.  Aerosol delivery via invasive ventilation: a narrative review.

Authors:  Hui-Ling Lin; James B Fink; Huiqing Ge
Journal:  Ann Transl Med       Date:  2021-04

2.  Practice pattern of aerosol therapy among patients undergoing mechanical ventilation in mainland China: A web-based survey involving 447 hospitals.

Authors:  Zhongheng Zhang; Peifeng Xu; Qiang Fang; Penglin Ma; Huiling Lin; Jim B Fink; Zongan Liang; Rongchang Chen; Huiqing Ge
Journal:  PLoS One       Date:  2019-08-29       Impact factor: 3.240

3.  The Clinical Practice and Best Aerosol Delivery Location in Intubated and Mechanically Ventilated Patients: A Randomized Clinical Trial.

Authors:  Chuanlin Zhang; Jie Mi; Zeju Zhang; Xueqin Wang; Yunxiao Zhu; Xinyi Luo; Ruiying Gan; Xiaoya Chen; Yujun Zou
Journal:  Biomed Res Int       Date:  2021-04-03       Impact factor: 3.411

Review 4.  Aerosol delivery during invasive mechanical ventilation: a systematic review.

Authors:  Jonathan Dugernier; Stephan Ehrmann; Thierry Sottiaux; Jean Roeseler; Xavier Wittebole; Thierry Dugernier; François Jamar; Pierre-François Laterre; Gregory Reychler
Journal:  Crit Care       Date:  2017-10-21       Impact factor: 9.097

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.