Use of the StarClose device for closure of inadvertent subclavian artery punctures.

Victor Tran1, Alexander Shiferson, Anil P Hingorani, Enrico Ascher, Natalie Marks, Kapil Gopal, Nirav Patel, Theresa Jacob.   

Abstract

PURPOSE: The placement of central catheters is a common procedure. It is also associated with multiple known complications. One of the potential complications that carry high morbidity and morality is arterial puncture and cannulation. Herein, we describe five case reports of a central line that was inadvertently placed in the subclavian artery and successfully removed using a StarClose device (Abbott Laboratories, Redwood CA). METHODS/
RESULTS: A retrospective chart review of a prospectively maintained database was performed. We identified five cases of inadvertent subclavian artery cannulation during central venous catheter placement. All catheters were removed successfully either in the operating room under fluoroscopic guidance or at the bedside with closure of the arteriotomy using the StarClose device. No cases required conversion to an open procedure for repair. No postremoval hematomas, bleeding episodes, myocardial infarctions, arrhythmias, or adverse clinical sequelae were identified. DISCUSSION: Based on our limited experience, we feel that this method can be performed safely and expeditiously not only in the operating room but also at the bedside.

Entities:  

Mesh:

Year:  2009        PMID: 19747613     DOI: 10.1016/j.avsg.2009.06.005

Source DB:  PubMed          Journal:  Ann Vasc Surg        ISSN: 0890-5096            Impact factor:   1.466


  2 in total

Review 1.  Closure Devices for Iatrogenic Thoraco-Cervical Vascular Injuries.

Authors:  Gregory C Makris; Rafiuddin Patel; Mark Little; Carina Tyrrell; James Sutcliffe; Kader Allouni; Mark Bratby; Susan Anthony; Raman Uberoi
Journal:  Cardiovasc Intervent Radiol       Date:  2016-11-28       Impact factor: 2.740

2.  Mini-invasive treatment of a large pseudoaneurysm of the neck related to central venous catheter placement: A case report.

Authors:  Chiara Palermo; Angelo Sanfiorenzo; Alessia Testo Giaquinta; Carla Virgilo; Massimiliano Veroux; Pierfrancesco Veroux
Journal:  Medicine (Baltimore)       Date:  2018-07       Impact factor: 1.889