Assessment of the clinical utility of an ultrasonic monitor of cardiac output (the USCOM) and agreement with thermodilution measurement.

Martin Boyle1, Liz Steel, Gordon M Flynn, Margherita Murgo, Lisa Nicholson, Maureen O'Brien, David Bihari.   

Abstract

OBJECTIVE: To assess the clinical utility of an ultrasonic monitor of cardiac output (USCOM), its reliability in tracking cardiac output (CO) changes and agreement with thermodilution (TD) measurements of CO.
DESIGN: Prospective comparison study. SETTING AND PARTICIPANTS: 55 adults undergoing thermodilution (TD) CO monitoring in a cardiothoracic or general intensive care unit between December 2006 and December 2007. MAIN OUTCOME MEASURES: USCOM and TD measurements of CO on two occasions in each patient were compared by Bland-Altman analysis for bias and limit of agreement. A mean percentage error <30% was considered acceptable. Per cent change in cardiac index (CI) was determined by each method. Doppler profiles obtained by the USCOM were assessed against an ideal standard ("acceptable").
RESULTS: 55 patients had measurements on 110 occasions, but Doppler waveforms were not obtained on 18 of these (16%), leaving 39 patients with paired comparisons for analysis (including 27 men; mean age, 64.7 [SD, 14.5] years). Mean TD CI was 3.4 +/-1.0L/min/m(2) (range, 2.0-6.0L/min/m(2)). The bias was 0.6L/min/m(2) (95% confidence limits [CLs], 0.4-0.8 L/min/m(2)), and the mean percentage error was 56% (95% CLs, 45%-65%). Twenty-two Doppler profiles (28%) were classed as acceptable; the mean percentage error for these was 62% (95% CLs, 38%-65%). On 15/19 occasions (74%) where TD CI changed > 15%, USCOM CI also changed >15%, but three of these changes (16%) were in the opposite direction. USCOM CI changed >15% on 9/20 occasions (45%) when TD CI did not.
CONCLUSIONS: Poor agreement with TD and a substantial rate of failure to obtain an USCOM measurement suggest that this device is unsuitable as a monitoring tool in intensive care.

Entities:  

Mesh:

Year:  2009        PMID: 19737122

Source DB:  PubMed          Journal:  Crit Care Resusc        ISSN: 1441-2772            Impact factor:   2.159


  4 in total

1.  Validation of electrical velocimetry in resuscitation of patients undergoing liver transplantation. Observational study.

Authors:  Ahmed M Mukhtar; Mohamed Elayashy; Amr H Sayed; Gihan M Obaya; Akram A Eladawy; Mai A Ali; Hisham M Dahab; Dina Z Khalaf; Mostafa A Mohamed; Amr H Elfouly; Gad M Behairy; Amr A Abdelaal
Journal:  J Clin Monit Comput       Date:  2019-04-19       Impact factor: 2.502

2.  Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM): An Ovine Cardiac Output Validation.

Authors:  Robert A Phillips; Sally G Hood; Beverley M Jacobson; Malcolm J West; Li Wan; Clive N May
Journal:  Crit Care Res Pract       Date:  2012-05-09

Review 3.  Hemodynamic monitoring in the critically ill: an overview of current cardiac output monitoring methods.

Authors:  Johan Huygh; Yannick Peeters; Jelle Bernards; Manu L N G Malbrain
Journal:  F1000Res       Date:  2016-12-16

Review 4.  Non-Invasive Monitoring of Cardiac Output in Critical Care Medicine.

Authors:  Lee S Nguyen; Pierre Squara
Journal:  Front Med (Lausanne)       Date:  2017-11-20
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.