Effect of audio coaching on correlation of abdominal displacement with lung tumor motion.

Mitsuhiro Nakamura1, Yuichiro Narita, Yukinori Matsuo, Masaru Narabayashi, Manabu Nakata, Akira Sawada, Takashi Mizowaki, Yasushi Nagata, Masahiro Hiraoka.   

Abstract

PURPOSE: To assess the effect of audio coaching on the time-dependent behavior of the correlation between abdominal motion and lung tumor motion and the corresponding lung tumor position mismatches. METHODS AND MATERIALS: Six patients who had a lung tumor with a motion range >8 mm were enrolled in the present study. Breathing-synchronized fluoroscopy was performed initially without audio coaching, followed by fluoroscopy with recorded audio coaching for multiple days. Two different measurements, anteroposterior abdominal displacement using the real-time positioning management system and superoinferior (SI) lung tumor motion by X-ray fluoroscopy, were performed simultaneously. Their sequential images were recorded using one display system. The lung tumor position was automatically detected with a template matching technique. The relationship between the abdominal and lung tumor motion was analyzed with and without audio coaching.
RESULTS: The mean SI tumor displacement was 10.4 mm without audio coaching and increased to 23.0 mm with audio coaching (p < .01). The correlation coefficients ranged from 0.89 to 0.97 with free breathing. Applying audio coaching, the correlation coefficients improved significantly (range, 0.93-0.99; p < .01), and the SI lung tumor position mismatches became larger in 75% of all sessions.
CONCLUSION: Audio coaching served to increase the degree of correlation and make it more reproducible. In addition, the phase shifts between tumor motion and abdominal displacement were improved; however, all patients breathed more deeply, and the SI lung tumor position mismatches became slightly larger with audio coaching than without audio coaching.

Entities:  

Mesh:

Year:  2009        PMID: 19735881     DOI: 10.1016/j.ijrobp.2008.11.070

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology.

Authors:  Edward D Brandner; Indrin J Chetty; Tawfik G Giaddui; Ying Xiao; M Saiful Huq
Journal:  Med Phys       Date:  2017-04-20       Impact factor: 4.071

2.  Effects of audio coaching and visual feedback on the stability of respiration during radiotherapy.

Authors:  Fumiya Baba; Satoshi Tanaka; Yoshinori Nonogaki; Shinji Hasegawa; Minami Nishihashi; Shiho Ayakawa; Maho Yamada; Yuta Shibamoto
Journal:  Jpn J Radiol       Date:  2016-06-17       Impact factor: 2.374

3.  Technical Note: Evaluation of audiovisual biofeedback smartphone application for respiratory monitoring in radiation oncology.

Authors:  Dante P I Capaldi; Tomi F Nano; Hao Zhang; Lawrie B Skinner; Lei Xing
Journal:  Med Phys       Date:  2020-10-10       Impact factor: 4.071

4.  Minimizing dose variation from the interplay effect in stereotactic radiation therapy using volumetric modulated arc therapy for lung cancer.

Authors:  Kazuki Kubo; Hajime Monzen; Mikoto Tamura; Makoto Hirata; Kentaro Ishii; Wataru Okada; Ryuta Nakahara; Shun Kishimoto; Ryu Kawamorita; Yasumasa Nishimura
Journal:  J Appl Clin Med Phys       Date:  2018-01-25       Impact factor: 2.102

5.  Assessment of tumor motion reproducibility with audio-visual coaching through successive 4D CT sessions.

Authors:  Samuel Goossens; Frédéric Senny; John A Lee; Guillaume Janssens; Xavier Geets
Journal:  J Appl Clin Med Phys       Date:  2014-01-04       Impact factor: 2.102

6.  Effect of audio instruction on tracking errors using a four-dimensional image-guided radiotherapy system.

Authors:  Mitsuhiro Nakamura; Akira Sawada; Nobutaka Mukumoto; Kunio Takahashi; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.