Literature DB >> 19735475

Figulla ASD occluder versus Amplatzer Septal Occluder: a comparative study on validation of a novel device for percutaneous closure of atrial septal defects.

Aysenur Pac1, Tugcin Bora Polat, Ilker Cetin, Mehmet Burhan Oflaz, Sevket Balli.   

Abstract

OBJECTIVES: Occlutech Figulla ASD Occluder (FSO) is an alternative device to Amplatzer Septal Occluder (ASO) with some structural innovations including increased flexibility, minimizing the amount of material implanted, and absence of the left atrial clamp. We aimed to report our experiences with FSO and compare the outcomes of this novel device versus ASO.
INTERVENTIONS: Between December 2005 and February 2009, 75 patients diagnosed with secundum atrial septal defects underwent transcatheter closure. The FSO device was used in 33 patients, and the ASO was used in 42.
RESULTS: Patient characteristics, stretch size of the defect, device left disc size, procedure, and fluoroscopy time were similar between the groups. However, the difference between device waist size and stretched diameter of the defect was significantly higher, and device delivery sheath was significantly larger in FSO group and device left disc size was significantly lower in the FSO group. In all subjects, the residual shunt was small to trivial during follow-up and the reduction in prevalence of residual shunt with time was similar in both groups (P = 0.68). We found no differences in complication rate between the two devices; however, device embolization to the pulmonary bifurcation in one patient was recorded as major complication in FSO device group.
CONCLUSIONS: Both devices are clinically safe and effective in ASD closure. FSO device has similar outcomes when compared to ASO device. Difficulties in selecting the correct device size in larger defects and larger venous sheath requirement need to be evaluated in further studies.

Entities:  

Mesh:

Year:  2009        PMID: 19735475     DOI: 10.1111/j.1540-8183.2009.00497.x

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  11 in total

1.  Transthoracic echocardiographic guidance during transcatheter closure of atrial septal defects in children and adults.

Authors:  Abdullah Erdem; Turkay Sarıtas; Cenap Zeybek; Ilker Kemal Yucel; Nurdan Erol; Halil Demır; Ender Odemıs; Yalim Yalcın; Ahmet Celebı
Journal:  Int J Cardiovasc Imaging       Date:  2011-08-11       Impact factor: 2.357

2.  Comparison of Figulla Flex® and Amplatzer™ devices for atrial septal defect closure: A meta-analysis.

Authors:  Alvaro Aparisi; Roman J Arnold; Hipólito Gutiérrez; Ana Revilla; Ana Serrador; Benigno Ramos; Tania Rodriguez-Gabella; Alberto Campo; Carlos Baladrón; Itziar Gómez; Manuel Carrasco-Moraleja; José A San Roman; Ignacio J Amat-Santos
Journal:  Cardiol J       Date:  2020-04-24       Impact factor: 2.737

3.  Percutaneous closure of secundum type atrial septal defects: More than 5-year follow-up.

Authors:  Roel Jr Snijder; Maarten J Suttorp; Jurriën M Ten Berg; Martijn C Post
Journal:  World J Cardiol       Date:  2015-03-26

4.  Comparison of the Occlutech ® Figulla ® septal occluder and Amplatzer ® septal occluder for atrial septal defect device closure.

Authors:  Supaporn Roymanee; Worakan Promphan; Nakharin Tonklang; Kanjarut Wongwaitaweewong
Journal:  Pediatr Cardiol       Date:  2015-01-30       Impact factor: 1.655

5.  Transcatheter atrial septal defect closure using occlutech figulla device: a two-center experience.

Authors:  Hojat Mortezaeian; Keyhan Sayadpour Zanjani; Elaheh Malakan Rad
Journal:  J Tehran Heart Cent       Date:  2013-10-28

6.  Transesophageal echocardiography and fluoroscopy for percutaneous closure of atrial septal defects: A comparative study.

Authors:  Weize Xu; Jianhua Li; Jingjing Ye; Jin Yu; Jiangen Yu; Zewei Zhang
Journal:  Medicine (Baltimore)       Date:  2018-10       Impact factor: 1.817

7.  Preliminary Experience With the New Amplatzer™ Trevisio™ Delivery System in Transcatheter Atrial Septal Defect Closures in Children.

Authors:  Raymond N Haddad; Diala Khraiche; Damien Bonnet; Mathilde Meot; Sophie Malekzadeh-Milani
Journal:  Front Pediatr       Date:  2021-03-11       Impact factor: 3.418

8.  Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions.

Authors:  Masood Sadiq; Asif Ur Rehman; Amjad Mehmood
Journal:  Ann Pediatr Cardiol       Date:  2016 May-Aug

9.  Percutaneous Atrial Septal Defect Closure Using the Occlutech Figulla Device in Adults: More than 800 Patient-Years of Follow-Up.

Authors:  R J R Snijder; L E Renes; D Bosshardt; M J Suttorp; J M Ten Berg; M C Post
Journal:  J Interv Cardiol       Date:  2020-02-14       Impact factor: 2.279

10.  Microtransesophageal Echocardiographic Guidance during Percutaneous Interatrial Septal Closure without General Anaesthesia.

Authors:  Roel J R Snijder; Laura E Renes; Martin J Swaans; Maarten Jan Suttorp; Jurrien M Ten Berg; Martijn C Post
Journal:  J Interv Cardiol       Date:  2020-09-07       Impact factor: 2.279

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.