Acute failure of a St. Jude's prosthetic aortic valve: large pannus formation masked by a small thrombus.

Seth Eric Hurwitz1, Daniel Waxman, Susan Hecht.   

Abstract

Pannus formation and valve thrombus can cause prosthetic valve failure. The authors report the case of a 50-year-old woman who presented to the emergency room with decompensated heart failure secondary to mechanical valve dysfunction. On two-dimensional and transesophageal echocardiography, the patient had severe aortic stenosis and regurgitation. A thrombus seen on the valve was felt to be the etiology of her prosthetic valve failure. She underwent emergent cardiac surgery for aortic valve replacement. Pathology revealed that although a small thrombus was present, extensive pannus was the underlying mechanism of valve dysfunction. Differentiation between pannus and thrombus may have important clinical implications, but this case illustrates that distinguishing between these entities by echocardiographic and clinical criteria may not be possible.

Entities:  

Mesh:

Year:  2009        PMID: 19733796     DOI: 10.1016/j.echo.2009.04.001

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  5 in total

1.  Cardiac masses : Experience from a Turkish tertiary center of cardiology.

Authors:  C Gecmen; G G Gecmen; M Kahyaoglu; B Omar; S Izci; A Kalayci; C Y Karabay; S Coban; O Candan; E Yanik; I A Izgi; N O Barisik
Journal:  Herz       Date:  2016-12-01       Impact factor: 1.443

2.  Acute obstruction of a mechanical aortic valve in a young woman: case report and review of the literature.

Authors:  Ashraf Roshdy; Georgios T Karapanagiotidis; Mazin A I Sarsam; Simon N Fletcher
Journal:  Echo Res Pract       Date:  2014-12-10

3.  Rare early prosthesis obstruction after mitral valve replacement: a case report and literature review.

Authors:  Jun Shi; Zhi-xuan Bai; Jia Hu; Ben-gui Zhang; Ying-qiang Guo
Journal:  J Cardiothorac Surg       Date:  2012-07-02       Impact factor: 1.637

4.  Artificial aortic valve dysfunction due to pannus and thrombus - different methods of cardiac surgical management.

Authors:  Stanisław Ostrowski; Anna Marcinkiewicz; Anna Kośmider; Andrzej Walczak; Radosław Zwoliński; Ryszard Jaszewski
Journal:  Kardiochir Torakochirurgia Pol       Date:  2015-09-28

5.  Effects of pannus formation on the flow around a bileaflet mechanical heart valve.

Authors:  Woojin Kim; Haecheon Choi; Jihoon Kweon; Dong Hyun Yang; Young-Hak Kim
Journal:  PLoS One       Date:  2020-06-12       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.