Literature DB >> 19721315

[Influence on dose calculation by difference of dose calculation algorithms in stereotactic lung irradiation: comparison of pencil beam convolution (inhomogeneity correction: batho power law) and analytical anisotropic algorithm].

Masayuki Tachibana1, Yoshitaka Noguchi, Jyunichi Fukunaga, Naomi Hirano, Satoshi Yoshidome, Takaaki Hirose.   

Abstract

The monitor unit (MU) was calculated by pencil beam convolution (inhomogeneity correction algorithm: batho power law) [PBC (BPL)] which is the dose calculation algorithm based on measurement in the past in the stereotactic lung irradiation study. The recalculation was done by analytical anisotropic algorithm (AAA), which is the dose calculation algorithm based on theory data. The MU calculated by PBC (BPL) and AAA was compared for each field. In the result of the comparison of 1031 fields in 136 cases, the MU calculated by PBC (BPL) was about 2% smaller than that calculated by AAA. This depends on whether one does the calculation concerning the extension of the second electrons. In particular, the difference in the MU is influenced by the X-ray energy. With the same X-ray energy, when the irradiation field size is small, the lung pass length is long, the lung pass length percentage is large, and the CT value of the lung is low, and the difference of MU is increased.

Mesh:

Year:  2009        PMID: 19721315     DOI: 10.6009/jjrt.65.1064

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  3 in total

1.  Impact of tissue heterogeneity corrections in stereotactic body radiation therapy treatment plans for lung cancer.

Authors:  Tania De La Fuente Herman; Heather Gabrish; Terence S Herman; Maria T Vlachaki; Salahuddin Ahmad
Journal:  J Med Phys       Date:  2010-07

2.  Stereotactic body radiation therapy (SBRT) and respiratory gating in lung cancer: dosimetric and radiobiological considerations.

Authors:  Tania De La Fuente Herman; Maria T Vlachaki; Terence S Herman; Kerry Hibbitts; Julie A Stoner; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2010-01-29       Impact factor: 2.102

3.  Evaluation of Lung Density and Its Dosimetric Impact on Lung Cancer Radiotherapy: A Simulation Study.

Authors:  T Raj Verma; N Kumar Painuly; S Prasad Mishra; S A Yoganathan; N Singh; M L B Bhatt; N Jamal
Journal:  J Biomed Phys Eng       Date:  2019-02-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.