Literature DB >> 19719488

Management strategies when implanted cardioverter defibrillator leads fail: survey findings.

Wenjie Xu1, Hans J Moore, Pamela E Karasik, Michael R Franz, Steve N Singh, Ross D Fletcher.   

Abstract

BACKGROUND: Defibrillator implanters have adopted different approaches to managing failures of multicomponent implanted cardioverter defibrillator (ICD) leads. Although recent publications identified single-component failures as common mechanisms of failure, there are no published data regarding how best to manage these failures.
METHODS: An internet-based survey was conducted to identify current management strategies. Questions were asked regarding isolated failure of a high-voltage coil or of a pace/sense electrode, in order to identify the frequency of various techniques to correct these failures.
RESULTS: A worldwide query collected strategies from 376 physicians identifying themselves as ICD-implanting physicians. Replies came from 28 countries, with the USA accounting for 83.2%. The survey was completed by 85.6% of respondents. Implant experience was >10 years for 61.1%, 3-10 years for 29.1%, and <3 years for 10.4%. When the right ventricular coil failed, 52% abandoned and 48% explanted the failed lead. In superior vena cava coil failure, 61.2% chose to simply exclude this coil, using the other intact lead components. For pace/sense defects, 53.1% chose to implant a new pace/sense lead or switch sensing electrodes, using the intact lead components. Medical literature (76.1%), personal experience (67.6%), and professional guidelines (63.7%) were strong decision-making influences.
CONCLUSIONS: (1) Management decisions for single-component failures of ICD leads are complex; (2) Significant differences in management strategy exist among physicians; (3) Medical literature and professional guidelines are strong influences for these decisions; (4) A lead failure registry could help identify reasons for such differences and help guide management.

Entities:  

Mesh:

Year:  2009        PMID: 19719488     DOI: 10.1111/j.1540-8159.2009.02454.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  The effect of device advisories on implantable cardioverter-defibrillator therapy.

Authors:  Mitesh S Amin; Kenneth A Ellenbogen
Journal:  Curr Cardiol Rep       Date:  2010-09       Impact factor: 2.931

2.  Spanish Rhythm Association member´s perspectives on cardiac implantable electronic device reuse in low- and middle-income countries.

Authors:  Iñigo Lorenzo Ruiz; Larraitz Gaztañaga Arantzamendi; Xabier Marichalar Mendia
Journal:  J Interv Card Electrophysiol       Date:  2022-07-15       Impact factor: 1.759

3.  Reuse of a 20-year-old lead: An abandoned lead reactivated.

Authors:  Breno Bernardes de Souza; Noel G Boyle
Journal:  HeartRhythm Case Rep       Date:  2015-08-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.