In vitro evaluation of the type of implant bed preparation with osteotomes in bone type IV and its influence on the stability of two implant systems.

Nuria García-Vives1, Rodrigo Andrés-García, Vicente Rios-Santos, Ana Fernández-Palacín, Pedro Bullón-Fernández, Mariano Herrero-Climent, Federico Herrero-Climent.   

Abstract

OBJECTIVES: 1) To evaluate and compare the stability of two types of implants in an animal model after preparing the implant bed with various sequences of osteotomes in bone type IV. 2) To evaluate the hypothesis of whether or not the sequence of using osteotomes influences the primary implant stability. MATERIAL AND
METHOD: We selected bone from cow ribs, which in its cross section (the most medullar area of the bone) would be equivalent to a type IV human bone. We used fifteen ribs, in which six implant beds were prepared in each rib block using different preparation protocols for seating three conical Swiss Plus SPB implants of 3.7 x 10 mm versus three MK III parallel wall implants of 4 x 10 mm.Three preparations, each with a diameter, were made for the implants, using osteotomes that progressively increased the diameter of the implant bed. In the first preparation, the complete sequence of osteotomes was used; in the second preparation, the last osteotome was left out; and in the third preparation, the implant was placed after only passing through the first osteotome. Once the implants were seated, we proceeded to evaluate the stability (Osstell ISQ-value). The locations were randomly chosen (by a coin toss). RESULTS AND STATISTICAL ANALYSIS: We performed a statistical analysis of the ISQ values that were obtained during the different preparations carried out for the Mk III and Swiss Plus implants. The average range and standard deviation were calculated. The hypothesis was compared by a two-way variance analysis (type of implant/ different sequences of preparing the implant bed). It was considered significant for a p <0.05. The statistical results obtained for the values of the Mk III implant were significant (p = 0.042).
CONCLUSIONS: The results of this in vitro study conclude that the tapered shape provides more primary stability to the implant and suggest that a short sequence of osteotomes in bone type IV provide more primary stability than the complete sequence.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19718009

Source DB:  PubMed          Journal:  Med Oral Patol Oral Cir Bucal        ISSN: 1698-4447


  16 in total

1.  Assessment of Osstell ISQ's reliability for implant stability measurement: a cross-sectional clinical study.

Authors:  Mariano Herrero-Climent; Rocio Santos-García; Reyes Jaramillo-Santos; Manuel-Maria Romero-Ruiz; Ana Fernández-Palacin; Pedro Lázaro-Calvo; Pedro Bullón; Jose-Vicente Ríos-Santos
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-11-01

2.  The effect of shape, length and diameter of implants on primary stability based on resonance frequency analysis.

Authors:  Hamidreza Barikani; Shadab Rashtak; Soolmaz Akbari; Mohammadjavad Kharrazi Fard; Amirreza Rokn
Journal:  Dent Res J (Isfahan)       Date:  2014-01

3.  Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis.

Authors:  Gorka Santamaría-Arrieta; Aritza Brizuela-Velasco; Felipe J Fernández-González; David Chávarri-Prado; Yelko Chento-Valiente; Eneko Solaberrieta; Markel Diéguez-Pereira; José-Antonio Vega; Jaime Yurrebaso-Asúa
Journal:  J Clin Exp Dent       Date:  2016-07-01

4.  The influence of length of implant on primary stability: An in vitro study using resonance frequency analysis.

Authors:  Anwar B Bataineh; Ala M Al-Dakes
Journal:  J Clin Exp Dent       Date:  2017-01-01

5.  In vitro comparison of primary stability of two implant designs in D3 bone.

Authors:  J González-Serrano; R Ortega-Aranegui; J López-Quiles
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2017-07-01

6.  Heat Generation at the Implant-Bone Interface by Insertion of Ceramic and Titanium Implants.

Authors:  Holger Zipprich; Paul Weigl; Eugenie König; Alexandra Toderas; Ümniye Balaban; Christoph Ratka
Journal:  J Clin Med       Date:  2019-09-25       Impact factor: 4.241

7.  Retrospective clinical study on sinus bone graft and tapered-body implant placement.

Authors:  Jong-Hwa Kim; Young-Kyun Kim; Ji-Hyun Bae
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2013-04-23

8.  The effect of implant length and diameter on the primary stability in different bone types.

Authors:  Hamidreza Barikani; Shadab Rashtak; Soolmaz Akbari; Samareh Badri; Niloufar Daneshparvar; Amirreza Rokn
Journal:  J Dent (Tehran)       Date:  2013-09-30

9.  Effect of implant macro-design on primary stability: A prospective clinical study.

Authors:  Naroa Lozano-Carrascal; Oscar Salomó-Coll; Marta Gilabert-Cerdà; Nuria Farré-Pagés; Jordi Gargallo-Albiol; Federico Hernández-Alfaro
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2016-03-01

10.  Influence of short implants geometry on primary stability.

Authors:  J González-Serrano; P Molinero-Mourelle; B Pardal-Peláez; L-M Sáez-Alcaide; R Ortega; J López-Quiles
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2018-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.