PURPOSE: To show the signal intensity varies with shot number in diffusion-weighted (DW) echo-planar imaging (EPI) and affects apparent diffusion coefficient (ADC) calculation.
MATERIALS AND METHODS: This prospective study was performed on 35 adult patients and 20 volunteers. Measurements were made on a 3T scanner using a breathhold DW spin-echo EPI (SE EPI) sequence. Three protocols were used: A) eight consecutive shots at a fixed b-value of 0 seconds/mm(2) with TR = 1000 and 3000 msec; B) seven consecutive shots at b-values = 0, 1000, 750, 500, 250, 100, 0 seconds/mm(2) (in that order) with TR = 3500 msec; and C) seven consecutive shots (as in B) with TR = 1000, 1750, and 7000 msec.
RESULTS: For protocol A, signal intensity decreased significantly from the first to second shot (P<0.0001) and thereafter remained constant. For protocol B, the ADC depended on which b = 0 seconds/mm(2) image was used. Using the first b = 0 seconds/mm(2), the mean ADC was 15% higher than using the second b = 0 seconds/mm(2) (P<0.0001). For protocol C, the difference between ADC using the first b = 0 seconds/mm(2) and the second b = 0 seconds/mm(2) decreased as the TR increased.
CONCLUSION: The signal intensity can vary with shot number in SE EPI. For TR > or = 3000 msec, steady-state is attained after one shot. Using data acquired prior to steady-state confounds the calculation of ADC values.

Entities:  

Mesh:

Year:  2009        PMID: 19711409     DOI: 10.1002/jmri.21861

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  3 in total

1.  Cardiac motion in diffusion-weighted MRI of the liver: artifact and a method of correction.

Authors:  Joy Liau; Jimmy Lee; Michael E Schroeder; Claude B Sirlin; Mark Bydder
Journal:  J Magn Reson Imaging       Date:  2011-09-29       Impact factor: 4.813

2.  Liver diffusivity in healthy volunteers and patients with chronic liver disease: comparison of breathhold and free-breathing techniques.

Authors:  Mamak Eatesam; Susan M Noworolski; Phyllis C Tien; Michelle Nystrom; Jennifer L Dodge; Raphael B Merriman; Aliya Qayyum
Journal:  J Magn Reson Imaging       Date:  2011-10-27       Impact factor: 4.813

3.  Dynamic characteristics of MR diffusion-weighted imaging in a rabbit liver VX-2 tumor model.

Authors:  You-Hong Yuan; En-Hua Xiao; Zhong He; Ke Jin; Cong Ma; Jun Xiang; Jian-Hua Xiao; Wei-Jian Chen
Journal:  Jpn J Radiol       Date:  2012-12-13       Impact factor: 2.374

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.