Comparison of the performance of four laryngoscopes in a high-fidelity simulator using normal and difficult airway.

L Powell1, J Andrzejowski, R Taylor, D Turnbull.   

Abstract

BACKGROUND: Novel laryngoscopes may demonstrate advantages over the traditional Macintosh laryngoscope for difficult airways. Our study compared Glidescope laryngoscope, Bonfils fibrescope, and CTrach intubating laryngeal mask airway with the Macintosh laryngoscope in a simulated normal and difficult airway, considering the additional effect of previous intubation experience.
METHODS: Twenty-two non-anaesthetists, 21 trainees, and 20 consultant anaesthetists attempted tracheal intubation of a Laerdal SimMan manikin, comparing a normal with a difficult airway scenario for each intubation device. The time taken to view the vocal cords and time to intubate were recorded. Also success rate and ease of use for each device were scored, alongside scope preference for each scenario.
RESULTS: Time to intubate was significantly shorter with the Macintosh compared with all three novel devices. All the devices had a high first-attempt success rate, but the Glidescope had a 100% first time successful intubation for all participants in both a normal and a difficult airway. Non-anaesthetists took significantly longer time to intubate compared with consultant anaesthetists, but there was no difference between trainee and consultant anaesthetists. Higher proportions of participants found the Glidescope easy to use, compared with other devices. For the normal airway, the Macintosh was the preferred device, but for the difficult airway, the Glidescope was favoured.
CONCLUSIONS: In this study, the Macintosh laryngoscope outperformed the other devices. However, the Glidescope was considered easy to use regardless of previous experience and was the preferred device for the simulated difficult airway.

Entities:  

Mesh:

Year:  2009        PMID: 19710072     DOI: 10.1093/bja/aep232

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  14 in total

Review 1.  [Indirect laryngoscopy : Alternatives to securing the airway].

Authors:  R R Noppens; C Werner; T Piepho
Journal:  Anaesthesist       Date:  2010-02       Impact factor: 1.041

2.  A comparison of the Trachway intubating stylet and the Macintosh laryngoscope in tracheal intubation: a manikin study.

Authors:  Kuo-Chuan Hung; Ping-Heng Tan; Victor Chia-Hsiang Lin; Hao-Kuang Wang; Hung-Shu Chen
Journal:  J Anesth       Date:  2012-10-07       Impact factor: 2.078

3.  [Direct laryngoscopy or C-MAC video laryngoscopy? Routine tracheal intubation in patients undergoing ENT surgery].

Authors:  D Meininger; U Strouhal; C F Weber; D Fogl; L Holzer; K Zacharowski; C Byhahn
Journal:  Anaesthesist       Date:  2010-08-13       Impact factor: 1.041

4.  A comparison of the C-MAC video laryngoscope to the Macintosh direct laryngoscope for intubation in the emergency department.

Authors:  John C Sakles; Jarrod Mosier; Stephen Chiu; Mari Cosentino; Leah Kalin
Journal:  Ann Emerg Med       Date:  2012-05-05       Impact factor: 5.721

5.  A comparative study on the usefulness of the Glidescope or Macintosh laryngoscope when intubating normal airways.

Authors:  Guen Seok Choi; Eun-Ha Lee; Chae Seong Lim; Seok-Hwa Yoon
Journal:  Korean J Anesthesiol       Date:  2011-05-31

6.  GlideScope Use improves intubation success rates: an observational study using propensity score matching.

Authors:  James W Ibinson; Catalin S Ezaru; Daniel S Cormican; Michael P Mangione
Journal:  BMC Anesthesiol       Date:  2014-11-05       Impact factor: 2.217

7.  Bonfils intubation fibrescope: use in simulation-based intubation training for medical students in comparison to MacIntosh laryngoscope.

Authors:  Tobias Limbach; Thomas Ott; Jan Griesinger; Antje Jahn-Eimermacher; Tim Piepho
Journal:  BMC Res Notes       Date:  2016-02-27

8.  Comparison of the Glidescope®, flexible fibreoptic intubating bronchoscope, iPhone modified bronchoscope, and the Macintosh laryngoscope in normal and difficult airways: a manikin study.

Authors:  Adrian Langley; Gabriel Mar Fan
Journal:  BMC Anesthesiol       Date:  2014-02-28       Impact factor: 2.217

9.  Comparison of GlideScope video laryngoscope with Macintosh laryngoscope in adult patients undergoing elective surgical procedures.

Authors:  Mrunalini Parasa; Srivishnu Vardhan Yallapragada; Nagendra Nath Vemuri; Mastan Saheb Shaik
Journal:  Anesth Essays Res       Date:  2016 May-Aug

10.  Intubation performance using different laryngoscopes while wearing chemical protective equipment: a manikin study.

Authors:  H Schröder; N Zoremba; R Rossaint; K Deusser; C Stoppe; M Coburn; A Rieg; G Schälte
Journal:  BMJ Open       Date:  2016-03-15       Impact factor: 2.692