Transcatheter valve insertion in a model of enlarged right ventricular outflow tracts.

Adeline Basquin1, Emmanuelle Pineau, Louise Galmiche, Damien Bonnet, Daniel Sidi, Younes Boudjemline.   

Abstract

OBJECTIVE: Transcatheter pulmonary valve insertion has recently emerged as an alternative to surgery. To extend its indications to patients with a large right ventricular outflow tract, we previously developed an intravascular device that reduces the diameter of the main pulmonary artery, allowing the insertion of available valved stents. Here we report its use in a model of animals with an enlarged right ventricular outflow tract and pulmonary valve incompetence. METHODS AND
RESULTS: The study comprised 33 sheep that first underwent surgical enlargement of the main pulmonary artery. We then intended to implant a filler percutaneously, followed later by the insertion of a valve. Three animals died during the intermediate stage. The remainder were humanely killed either immediately (group 1, n = 6) or after a mean follow-up of 1 (group 2, n = 12) or 2 months (group 3, n = 12). Animals from groups 2 and 3 were equally divided into 2 subgroups according to the difference between diameters of the device inserted and the main pulmonary artery (A < 5 mm, B > or = 5 mm). Fillers were all inserted successfully (n = 30), although one embolized after its insertion (group 3A). A valved stent was implanted in all animals, but in 1 case a balloon ruptured during inflation of the stent leading to incomplete expansion and the death of the animal. Six animals, 5 of which were from group A, had pulmonary regurgitation after valve insertion.
CONCLUSION: Pulmonary valve insertion is possible through a transcatheter technique using a pulmonary artery filler. Oversizing the device reduces the risk of embolization and paraprosthetic leak. Copyright 2010 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19709678     DOI: 10.1016/j.jtcvs.2009.07.025

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  3 in total

1.  An ovine model of pulmonary insufficiency and right ventricular outflow tract dilatation.

Authors:  J Daniel Robb; Matthew A Harris; Masahito Minakawa; Evelio Rodriguez; Kevin J Koomalsingh; Takashi Shuto; Yoav Dori; Robert C Gorman; Joseph H Gorman; Matthew J Gillespie
Journal:  J Heart Valve Dis       Date:  2012-03

Review 2.  Will catheter interventions replace surgery for valve abnormalities?

Authors:  Michael L O'Byrne; Matthew J Gillespie
Journal:  Curr Opin Cardiol       Date:  2014-01       Impact factor: 2.161

Review 3.  Tissue Engineered Transcatheter Pulmonary Valved Stent Implantation: Current State and Future Prospect.

Authors:  Xiling Zhang; Thomas Puehler; Jette Seiler; Stanislav N Gorb; Janarthanan Sathananthan; Stephanie Sellers; Assad Haneya; Jan-Hinnerk Hansen; Anselm Uebing; Oliver J Müller; Derk Frank; Georg Lutter
Journal:  Int J Mol Sci       Date:  2022-01-10       Impact factor: 5.923

  3 in total