Influence of repeated screw tightening on bacterial leakage along the implant-abutment interface.

Cássio do Nascimento1, Vinícius Pedrazzi, Paola Kirsten Miani, Larissa Daher Moreira, Rubens Ferreira de Albuquerque.   

Abstract

OBJECTIVES: Bacterial penetration along the implant-abutment interface as a consequence of abutment screw loosening has been reported in a number of recent studies. The aim of this in vitro study was to investigate the influence of repeated tightening of the abutment screw on leakage of Streptococcus mutans along the interface between implants and pre-machined abutments.
MATERIALS AND METHODS: Twenty pre-machined abutments with a plastic sleeve were used. The abutment screws were tightened to 32 N cm in group 1 (n=10 - control) and to 32 N cm, loosened and re-tightened with the same torque twice in group 2 (n=10). The assemblies were completely immersed in 5 ml of Tryptic Soy Broth medium inoculated with S. mutans and incubated for 14 days. After this period, contamination of the implant internal threaded chamber was evaluated using the DNA Checkerboard method.
RESULTS: Microorganisms were found on the internal surfaces of both groups evaluated. However, bacterial counts in group 2 were significantly higher than that in the control group (P<0.05).
CONCLUSION: These results suggest that bacterial leakage between implants and abutments occurs even under unloaded conditions and at a higher intensity when the abutment screw is tightened and loosened repeatedly.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19709063     DOI: 10.1111/j.1600-0501.2009.01769.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  6 in total

1.  Implant-abutment leaking of replace conical connection nobel biocare® implant system. An in vitro study of the microbiological penetration from external environment to implant-abutment space.

Authors:  E El Haddad; A B Giannì; G E Mancini; F Cura; F Carinci
Journal:  Oral Implantol (Rome)       Date:  2016-11-13

2.  Micro-leakage at the implant-abutment interface with different tightening torques in vitro.

Authors:  João Paulo da Silva-Neto; Marcel Santana Prudente; Thiago de Almeida Prado Naves Carneiro; Mauro Antônio de Arruda Nóbilo; Mario Paulo Amante Penatti; Flávio Domingues das Neves
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

3.  Impact of Dynamic Loading on the Implant-abutment Interface Using a Gas-enhanced Permeation Test In Vitro.

Authors:  Anas Al-Jadaa; Thomas Attin; Timo Peltomäki; Christian Heumann; Patrick Roger Schmidlin
Journal:  Open Dent J       Date:  2015-03-31

4.  In Vitro Evaluation of Bacterial Adhesion and Bacterial Viability of Streptococcus mutans, Streptococcus sanguinis, and Porphyromonas gingivalis on the Abutment Surface of Titanium and Zirconium Dental Implants.

Authors:  Ana Stefany Meza-Siccha; Miguel Angel Aguilar-Luis; Wilmer Silva-Caso; Fernando Mazulis; Carolina Barragan-Salazar; Juana Del Valle-Mendoza
Journal:  Int J Dent       Date:  2019-06-13

5.  In Vitro Bacterial Leakage at the Implant-Abutment Connection of Two Dental Implant Systems with Internal Connection.

Authors:  Azadeh Khajavi; Shabnam Mohseni; Amir Peymani; Mehrak Amjadi
Journal:  Front Dent       Date:  2020-12-08

6.  Biofilm and dental implant: The microbial link.

Authors:  Sangeeta Dhir
Journal:  J Indian Soc Periodontol       Date:  2013-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.