Ivana Lopes Romero1, João B N S Malta, Cely B Silva, Lycia M J Mimica, Kaz H Soong, Richard Y Hida.
Abstract
PURPOSE: To ascertain if the polymerization reaction also contributes additionally to the antibacterial effects of two commonly used cyanoacrylate tissue adhesives.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19700870 PMCID: PMC2804120 DOI: 10.4103/0301-4738.55065
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Figure 1(a) Müller-Hinton media plate cultured with Staphylococcus aureus. (b) Blood agar plate cultured with Streptococcus pneumoniae. Inhibition halos seen around the discs of ethyl-cyanoacrylate (EC) (black arrows) and N-butyl-cyanocrylate (BC) (white arrows). Central disc is control (gray arrow). (c) Measurement of inhibition halo with ruler
Mean and standard deviation (SD) of inhibitory halos (mm), comparing actively polymerizing liquid ethyl-cyanoacrylate discs (EC-liq) with previously polymerized ethyl-cyanoacrylate discs (EC-poly) for Staphylococcus aureus, Streptococcus pneumoniae, Escherichia coli, and Pseudomonas aeruginosa. Statistical significance of difference between EC-liq and EC-poly is given as P-values
| EC-liq | EC-poly | EC-liq | EC-poly | EC-liq | EC-poly | EC-liq | EC-poly | |
|---|---|---|---|---|---|---|---|---|
| Mean | 14.05 | 13.45 | 14.25 | 12.70 | 11.2 | 0 | 0 | 0 |
| SD | 1.00 | 1.00 | 1.33 | 1.38 | 2.67 | 0 | 0 | 0 |
| 0.019 | 0.0014 | <0.0001 | N/A | |||||
Mean and standard deviation (SD) of inhibitory halos (mm), comparing actively polymerizing liquid N-butylcyanoacrylate discs (BC-liq) with previously polymerized N-butyl-cyanoacrylate discs (BC-poly) for Staphylococcus aureus, Streptococcus pneumoniae, Escherichia coli, and Pseudomonas aeruginosa
| BC-liq | BC-poly | BC-liq | BC-poly | BC-liq | BC-poly | BC-liq | BC-poly | |
|---|---|---|---|---|---|---|---|---|
| Mean | 13.40 | 11.70 | 14.70 | 12.75 | 0 | 0 | 0 | 0 |
| SD | 1.05 | 0.86 | 2.13 | 1.25 | 0 | 0 | 0 | 0 |
| <0.0001 | 0.0010 | N/A | N/A | |||||
Statistical significance of difference between BC-liq and BC-poly is given as P-values.
Bactericidal effect of actively polymerizing liquid ethyl-cyanoacrylate (EC) and N-buthyl-cyanoacrylate (BC) (EC-liq and BC-liq) and previously polymerized (EC-poly and BC-poly) against the microorganisms in question
| EC-liq | EC-poly | BC-liq | BC-poly | |
|---|---|---|---|---|
| 20% | 10% | 0% | 10% | |
| 70% | 40% | 70% | 70% | |
| 30% | 0% | 0% | 0% | |
| 0% | 0% | 0% | 0% |