Literature DB >> 19693397

On-chip microfluidic systems for determination of L-glutamate based on enzymatic recycling of substrate.

W Laiwattanapaisal, J Yakovleva, M Bengtsson, T Laurell, S Wiyakrutta, V Meevootisom, O Chailapakul, J Emnéus.   

Abstract

Two microfluidic systems have been developed for specific analysis of L-glutamate in food based on substrate recycling fluorescence detection. L-glutamate dehydrogenase and a novel enzyme, D-phenylglycine aminotransferase, were covalently immobilized on (i) the surface of silicon microchips containing 32 porous flow channels of 235 mum depth and 25 mum width and (ii) polystyrene Poros beads with a particle size of 20 mum. The immobilized enzymes recycle L-glutamate by oxidation to 2-oxoglutarate followed by the transfer of an amino group from D-4-hydroxyphenylglycine to 2-oxoglutarate. The reaction was accompanied by reduction of nicotinamide adenine dinucleotide (NAD(+)) to NADH, which was monitored by fluorescence detection (epsilon(ex)=340 nm, epsilon(em)=460 nm). First, the microchip-based system, L-glutamate was detected within a range of 3.1-50.0 mM. Second, to be automatically determined, sequential injection analysis (SIA) with the bead-based system was investigated. The bead-based system was evaluated by both flow injection analysis and SIA modes, where good reproducibility for L-glutamate calibrations was obtained (relative standard deviation of 3.3% and 6.6%, respectively). In the case of SIA, the beads were introduced and removed from the microchip automatically. The immobilized beads could be stored in a 20% glycerol and 0.5 mM ethylenediaminetetraacetic acid solution maintained at a pH of 7.0 using a phosphate buffer for at least 15 days with 72% of the activity remaining. The bead-based system demonstrated high selectivity, where L-glutamate recoveries were between 91% and 108% in the presence of six other L-amino acids tested.

Entities:  

Year:  2009        PMID: 19693397      PMCID: PMC2717601          DOI: 10.1063/1.3098319

Source DB:  PubMed          Journal:  Biomicrofluidics        ISSN: 1932-1058            Impact factor:   2.800


  25 in total

1.  Lab-on-a-chip: a revolution in biological and medical sciences

Authors: 
Journal:  Anal Chem       Date:  2000-05-01       Impact factor: 6.986

2.  Covalent coupling methods for inorganic support materials.

Authors:  H H Weetall
Journal:  Methods Enzymol       Date:  1976       Impact factor: 1.600

Review 3.  Micro total analysis systems. Recent developments.

Authors:  Torsten Vilkner; Dirk Janasek; Andreas Manz
Journal:  Anal Chem       Date:  2004-06-15       Impact factor: 6.986

4.  Developments toward a microfluidic system for long-term monitoring of dynamic cellular events in immobilized human cells.

Authors:  Richard Davidsson; Ake Boketoft; Jesper Bristulf; Knut Kotarsky; Björn Olde; Christer Owman; Martin Bengtsson; Thomas Laurell; Jenny Emnéus
Journal:  Anal Chem       Date:  2004-08-15       Impact factor: 6.986

Review 5.  Lab-on-a-chip devices for global health: past studies and future opportunities.

Authors:  Curtis D Chin; Vincent Linder; Samuel K Sia
Journal:  Lab Chip       Date:  2006-10-27       Impact factor: 6.799

6.  Microchip immobilized enzyme reactors for hydrolysis of methyl cellulose.

Authors:  Claes Melander; Dane Momcilovic; Carina Nilsson; Martin Bengtsson; Herje Schagerlöf; Folke Tjerneld; Thomas Laurell; Curt T Reimann; Lo Gorton
Journal:  Anal Chem       Date:  2005-05-15       Impact factor: 6.986

7.  Single-channel microchip for fast screening and detailed identification of nitroaromatic explosives or organophosphate nerve agents.

Authors:  Joseph Wang; Martin Pumera; Madhu Prakash Chatrathi; Alberto Escarpa; Mustafa Musameh; Greg Collins; Ashok Mulchandani; Yuehe Lin; Khris Olsen
Journal:  Anal Chem       Date:  2002-03-01       Impact factor: 6.986

8.  Cholinomimetics increase glutamate outflow via an action on the corticostriatal pathway: implications for Alzheimer's disease.

Authors:  S N Dijk; P T Francis; G C Stratmann; D M Bowen
Journal:  J Neurochem       Date:  1995-11       Impact factor: 5.372

9.  Microfluidic biosensing systems. Part II. Monitoring the dynamic production of glucose and ethanol from microchip-immobilised yeast cells using enzymatic chemiluminescent micro-biosensors.

Authors:  Richard Davidsson; Bjorn Johansson; Volkmar Passoth; Martin Bengtsson; Thomas Laurell; Jenny Emneus
Journal:  Lab Chip       Date:  2004-09-14       Impact factor: 6.799

10.  Highly selective microbiosensors for in vivo measurement of glucose, lactate and glutamate.

Authors:  O M Schuvailo; O O Soldatkin; A Lefebvre; R Cespuglio; A P Soldatkin
Journal:  Anal Chim Acta       Date:  2006-03-16       Impact factor: 6.558

View more
  5 in total

1.  Organosilane deposition for microfluidic applications.

Authors:  Nick R Glass; Ricky Tjeung; Peggy Chan; Leslie Y Yeo; James R Friend
Journal:  Biomicrofluidics       Date:  2011-08-16       Impact factor: 2.800

2.  A perspective on microfluidic biofuel cells.

Authors:  Jin Wook Lee; Erik Kjeang
Journal:  Biomicrofluidics       Date:  2010-11-10       Impact factor: 2.800

Review 3.  Microfluidic opportunities in the field of nutrition.

Authors:  Sixing Li; Justin Kiehne; Lawrence I Sinoway; Craig E Cameron; Tony Jun Huang
Journal:  Lab Chip       Date:  2013-10-21       Impact factor: 6.799

Review 4.  Fish-on-a-chip: a sensitive detection microfluidic system for Alzheimer's disease.

Authors:  Jasmine P Devadhasan; Sanghyo Kim; Jeongho An
Journal:  J Biomed Sci       Date:  2011-05-28       Impact factor: 8.410

5.  Label Free Detection of L-Glutamate Using Microfluidic Based Thermal Biosensor.

Authors:  Varun Lingaiah Kopparthy; Siva Mahesh Tangutooru; Eric J Guilbeau
Journal:  Bioengineering (Basel)       Date:  2015-01-12
  5 in total