Literature DB >> 19690042

Combination of converging collimators for high-sensitivity brain SPECT.

Ruben Ter-Antonyan1, Ronald J Jaszczak, Kim L Greer, James E Bowsher, Scott D Metzler, R Edward Coleman.   

Abstract

UNLABELLED: The objective of this study, which is related to human brain SPECT, was to increase the sensitivity of a triple-camera SPECT system and reduce statistical noise in reconstructed images using a combination of converging collimators. The reason for combining collimators is to ensure both high sensitivity and sufficient sampling without trading off spatial resolution.
METHODS: A high-sensitivity half-cone-beam (HCB) collimator, designed specifically for brain imaging, was combined with other collimators and compared with conventional parallel-beam and fanbeam circular orbit acquisitions. For comparison, previously studied HCB collimation with a circle-and-helix data acquisition trajectory was also included in this study. Simulations of the Hoffman 3-dimensional brain phantom were performed to calculate the efficiencies of collimators and their combinations and to quantitatively evaluate reconstruction bias, statistical noise, and signal-to-noise ratios in the reconstructed images. Experimental brain phantom data were also acquired and compared for different acquisition types. Finally, a patient brain scan was obtained with a combination of HCB and fanbeam collimators and compared with a triple-fanbeam circular orbit acquisition.
RESULTS: A combination of 2 HCB collimators and 1 fanbeam collimator, compared with a triple-fanbeam collimator, can increase the photon detection efficiency by 27% and by more than a factor of 2, compared with triple-parallel-hole collimation, with equal spatial resolution measured on the axis of rotation. Quantitative analysis of reconstruction bias and visual analysis of the images showed no signs of sampling artifacts. Reconstructed images in the simulations, experimental brain phantom, and patient brain scans showed improved quality with this collimator combination due to increased sensitivity and reduced noise. Lesion visibility was also improved, as confirmed by signal-to-noise ratios. Alternatively, triple-HCB circle-and-helix acquisition has also shown competitive results, with a slight disadvantage in axial sampling and implementation procedure.
CONCLUSION: Combined HCB and fanbeam collimation is a promising approach for high-sensitivity brain SPECT.

Entities:  

Mesh:

Year:  2009        PMID: 19690042     DOI: 10.2967/jnumed.109.062653

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  3 in total

1.  Introduction of a novel ultrahigh sensitivity collimator for brain SPECT imaging.

Authors:  Mi-Ae Park; Marie Foley Kijewski; Ronnie Keijzers; Mark Keijzers; Morgan C Lyon; Laura Horky; Stephen C Moore
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

2.  Three-dimensional brain phantom containing bone and grey matter structures with a realistic head contour.

Authors:  Hidehiro Iida; Yuki Hori; Kenji Ishida; Etsuko Imabayashi; Hiroshi Matsuda; Masaaki Takahashi; Hirotaka Maruno; Akihide Yamamoto; Kazuhiro Koshino; Junichiro Enmi; Satoshi Iguchi; Tetsuaki Moriguchi; Hidekazu Kawashima; Tsutomu Zeniya
Journal:  Ann Nucl Med       Date:  2012-09-26       Impact factor: 2.668

3.  Fast quantitative reconstruction with focusing collimators for liver SPECT.

Authors:  Martijn M A Dietze; Sandra van der Velden; Marnix G E H Lam; Max A Viergever; Hugo W A M de Jong
Journal:  EJNMMI Phys       Date:  2018-12-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.