Literature DB >> 19686486

A comparison of stroke volume variation measured by the LiDCOplus and FloTrac-Vigileo system.

R B P de Wilde1, B F Geerts, P C M van den Berg, J R C Jansen.   

Abstract

The aim of this study was to compare the accuracy of stroke volume variation (SVV) as measured by the LiDCOplus system (SVVli) and by the FloTrac-Vigileo system (SVVed). We measured SVVli and SVVed in 15 postoperative cardiac surgical patients following five study interventions; a 50% increase in tidal volume, an increase of PEEP by 10 cm H2O, passive leg raising, a head-up tilt procedure and fluid loading. Between each intervention, baseline measurements were performed. 136 data pairs were obtained. SVVli ranged from 1.4% to 26.8% (mean (SD) 8.7 (4.6)%); SVVed from 2.0% to 26.0% (10.2 (4.7)%). The bias was found to be significantly different from zero at 1.5 (2.5)%, p < 0.001, (95% confidence interval 1.1-1.9). The upper and lower limits of agreement were found to be 6.4 and -3.5% respectively. The coefficient of variation for the differences between SVVli and SVVed was 26%. This results in a relative large range for the percentage limits of agreement of 52%. Analysis in repeated measures showed coefficients of variation of 21% for SVVli and 22% for SVVed. The LiDCOplus and FloTrac-Vigileo system are not interchangeable. Furthermore, the determination of SVVli and SVVed are too ambiguous, as can be concluded from the high values of the coefficient of variation for repeated measures. These findings underline Pinsky's warning of caution in the clinical use of SVV by pulse contour techniques.

Entities:  

Mesh:

Year:  2009        PMID: 19686486     DOI: 10.1111/j.1365-2044.2009.06009.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  4 in total

1.  Does intravenous landiolol, a β1-adrenergic blocker, affect stroke volume variation?

Authors:  Zen'ichiro Wajima; Toshiya Shiga; Kazuyuki Imanaga; Tetsuo Inoue
Journal:  J Anesth       Date:  2013-05-08       Impact factor: 2.078

2.  Effects of high-volume, rapid-fluid therapy on cardiovascular function and hematological values during isoflurane-induced hypotension in healthy dogs.

Authors:  Alexander Valverde; Giacomo Gianotti; Eva Rioja-Garcia; Amanda Hathway
Journal:  Can J Vet Res       Date:  2012-04       Impact factor: 1.310

Review 3.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

4.  Orthostatic intolerance after fast-track knee arthroplasty: Incidence and hemodynamic pathophysiology.

Authors:  Ana-Marija Hristovska; Louise B Andersen; Mette Grentoft; Jesper Mehlsen; Kirill Gromov; Henrik Kehlet; Nicolai B Foss
Journal:  Acta Anaesthesiol Scand       Date:  2022-07-14       Impact factor: 2.274