Literature DB >> 19684031

Reconstruction of atrioventricular valves with photo-oxidized bovine pericardium.

Peter Verbrugghe1, Bart Meuris, Willem Flameng, Paul Herijgers.   

Abstract

Repair of complex valve pathological processes often requires the use of leaflet tissue or pericardium. The use of bovine photo-oxidized pericardium may be an alternative, a tissue less prone to calcification. The aim of this study is to evaluate the use of photo-oxidized bovine pericardial tissue in the reconstruction of atrioventricular valves in humans. Between July 2001 and September 2006, 21 patients with complex valve pathology underwent a reconstruction with photo-oxidized pericardium. The pericardial patch was used for the reconstruction of a tricuspid valve leaflet in two patients, the reconstruction of a mitral valve leaflet in six patients, the reconstruction of the tricuspid annulus in one patient and the reconstruction of the mitral annulus in 12 patients. The follow-up ranged from 13.9 to 43.2 months. There were five perioperative deaths. Four patients developed failure of the reconstruction, in one patient there was thinning and perforation of the pericardial patch without signs of infection or abrasion. The other patients were free from thromboembolism, endocarditis, hemorrhagic complications or echocardiographic signs of failure of the reconstruction. Photo-oxidized bovine pericardium is a versatile material for complex reconstruction of the atrioventricular valvular structures. Its durability should, however, be investigated in comparison with alternative tissues in a randomized trial.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19684031     DOI: 10.1510/icvts.2008.200097

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  3 in total

1.  Decellularized aortic conduits: could their cryopreservation affect post-implantation outcomes? A morpho-functional study on porcine homografts.

Authors:  Michele Gallo; Antonella Bonetti; Helen Poser; Filippo Naso; Tomaso Bottio; Roberto Bianco; Adolfo Paolin; Paolo Franci; Roberto Busetto; Anna Chiara Frigo; Edward Buratto; Michele Spina; Maurizio Marchini; Fulvia Ortolani; Laura Iop; Gino Gerosa
Journal:  Heart Vessels       Date:  2016-04-26       Impact factor: 2.037

2.  The susceptibility of bioprosthetic heart valve leaflets to oxidation.

Authors:  Abigail J Christian; Hongqiao Lin; Ivan S Alferiev; Jeanne M Connolly; Giovanni Ferrari; Stanley L Hazen; Harry Ischiropoulos; Robert J Levy
Journal:  Biomaterials       Date:  2013-12-18       Impact factor: 12.479

3.  Vascular Remodeling of Clinically Used Patches and Decellularized Pericardial Matrices Recellularized with Autologous or Allogeneic Cells in a Porcine Carotid Artery Model.

Authors:  Jaroslav Chlupac; Roman Matejka; Miroslav Konarik; Robert Novotny; Zuzana Simunkova; Iveta Mrazova; Ondrej Fabian; Milan Zapletal; Zdenek Pulda; Jan Falk Lipensky; Jana Stepanovska; Karel Hanzalek; Antonin Broz; Tomas Novak; Alena Lodererova; Ludek Voska; Theodor Adla; Jiri Fronek; Miroslav Rozkot; Serhiy Forostyak; Peter Kneppo; Lucie Bacakova; Jan Pirk
Journal:  Int J Mol Sci       Date:  2022-03-18       Impact factor: 5.923