BACKGROUND: In 1997 the German Red Cross (GRC) blood donor services introduced mini-pool nucleic acid testing (NAT) for human immunodeficiency virus (HIV)-1, hepatitis C virus (HCV) and hepatitis B virus (HBV) to increase blood safety. With the new cobas s 201/cobas TaqScreen MPX, a fully automated extraction method and a multiplex amplification system specifically adapted to the needs of blood donation services is available.
METHODS: The cobas s 201 system was evaluated at the GRC BTS locations Hagen, Springe and Frankfurt. In phase A, the analytical sensitivity for the detection of HBV, HCV and HIV-1 was investigated and in phase B, at least 60,000 samples at each test site were screened in parallel with the MPX test on s 201 system and the existing routine mini-pool NAT system to compare the diagnostic specificity and the diagnostic sensitivity.
RESULTS: Comparable analytical sensitivities in a range of 1.6-3.6 IU/ml, 4.9-10.9 IU/ml and 14.7-26.6 IU/ml for HBV, HCV HIV, respectively, for the MPX test on s 201 system (95% probability based on probit analysis) were determined at all test sites. The diagnostic sensitivity was 99.8% and the diagnostic specificity was 99.85%.
CONCLUSIONS: The MPX test on s 201 system is a fully automated NAT system suitable for routine blood donor screening. The analytical sensitivity as well as the diagnostic sensitivity fulfilled all requirements of the Paul Ehrlich Institute for blood donor screening in mini-pools up to 96 donations per pool. A major benefit of the automated NAT system is the reduced personnel time and the extensive complete barcode-controlled process documentation.

Entities:  

Mesh:

Year:  2009        PMID: 19682348     DOI: 10.1111/j.1423-0410.2009.01219.x

Source DB:  PubMed          Journal:  Vox Sang        ISSN: 0042-9007            Impact factor:   2.144


  8 in total

1.  Validation of Virus NAT for HIV, HCV, HBV and HAV Using Post-Mortal Blood Samples.

Authors:  Knut Gubbe; Yvonne Scharnagl; Steffi Grosch; Torsten Tonn; Michael Schmidt; Kai M Hourfar; Andreas Karl; Erhard Seifried; Ina Wilkemeyer; Ulrich Kalus
Journal:  Transfus Med Hemother       Date:  2012-11-13       Impact factor: 3.747

2.  Design and Development of a Multiplex Real-Time PCR Assay for Detection of HIV-1 and HCV Using Molecular Beacons.

Authors:  Mahdi Paryan; Samira Mohammadi-Yeganeh; Siamak Mirab Samiee; Houri Rezvan
Journal:  Indian J Microbiol       Date:  2012-05-11       Impact factor: 2.461

3.  Prevalence of HIV Indeterminate Western Blot Tests and Follow-up of HIV Antibody Sero-Conversion in Southeastern China.

Authors:  Shouli Wu; Min Gao; Jian Zheng; Pingping Yan; Wei Wang; Xiaoli Lu; Yuefeng Qiu; Yansheng Yan
Journal:  Virol Sin       Date:  2019-06-12       Impact factor: 4.327

4.  Group testing case identification with biomarker information.

Authors:  Dewei Wang; Christopher S McMahan; Joshua M Tebbs; Christopher R Bilder
Journal:  Comput Stat Data Anal       Date:  2018-02-01       Impact factor: 1.681

Review 5.  Molecular mechanisms underlying occult hepatitis B virus infection.

Authors:  Jasmine Samal; Manish Kandpal; Perumal Vivekanandan
Journal:  Clin Microbiol Rev       Date:  2012-01       Impact factor: 26.132

6.  Estimating the prevalence of multiple diseases from two-stage hierarchical pooling.

Authors:  Md S Warasi; Joshua M Tebbs; Christopher S McMahan; Christopher R Bilder
Journal:  Stat Med       Date:  2016-04-18       Impact factor: 2.373

7.  Array testing for multiplex assays.

Authors:  Peijie Hou; Joshua M Tebbs; Dewei Wang; Christopher S McMahan; Christopher R Bilder
Journal:  Biostatistics       Date:  2020-07-01       Impact factor: 5.899

8.  Nucleic acid testing: Is it the only answer for safe Blood in India?

Authors:  N K Naidu; Z S Bharucha; Vandana Sonawane; Imran Ahmed
Journal:  Asian J Transfus Sci       Date:  2016 Jan-Jun
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.