| Literature DB >> 19675784 |
Simon Pridgeon1, Christopher Harding, Douglas Newton, Robert Pickard.
Abstract
OBJECTIVE: To evaluate the accuracy and diagnostic usefulness of a disposable flowmeter consisting of a plastic funnel with a spout divided into three chambers.Entities:
Keywords: Bladder outlet obstruction; urinary symptoms; uroflowmetry
Year: 2007 PMID: 19675784 PMCID: PMC2721516 DOI: 10.4103/0970-1591.32058
Source DB: PubMed Journal: Indian J Urol ISSN: 0970-1591
Figure 1Photographs of the device showing a, the 3-chamber design and b, the 4.6 mm aperture
Figure 2Scatter plot showing the relationship between maximum flow rate (Qmax) readings obtained in the office using the standard uroflowmetry (vertical axis) and those obtained with the funnel device (horizontal axis). The mean of the single standard office measurements for subjects voiding within each funnel category is indicated by the horizontal bar
Accuracy of device in randomized clinic-based comparison with standard uroflowmetry according to Qmax thresholds
| Funnel category | Standard uroflowmetry Qmax (ml s−1) | Predictive value (%) | |||
|---|---|---|---|---|---|
| < 10 | 10 - 15 | > 15 | PPV | NPV | |
| Bottom (<10 ml s−1) | 8 | 3 | 1 | 70 | 94 |
| Middle (10 - 15 s−1) | 2 | 10 | 8 | 50 | 88 |
| Top/cup (>15 ml s−1) | 0 | 0 | 14 | 100 | 72 |
PPV = Positive predictive value; NPV = Negative predictive value
Figure 3X-Y plot of the averaged notional home Qmax (horizontal axis) for 14 men against their single office-based standard reading (vertical axis). The broken lines show a diagnostic cut-off value of 15 ml s-1. Note plots for 12 of 14 (86%) subjects lie in the upper left or lower right quadrants indicating good agreement.
Comparison of the difference in voided volume (Vvoid) and the degree of error in estimation of standard single office Qmax by home use of the funnel device
| Degree of error of home readings using funnel device compared to single office Qmax (number of funnel categories) | |||||
|---|---|---|---|---|---|
| +2 | +1 | 0 | −1 | −2 | |
| Number of voids | 0 | 3 | 124 | 49 | 20 |
| Mean (SD) difference in Vvoid(ml) | −78 (185) | −17 (113) | −68 (139) | −165 (109) | |
Single office flow was assigned to appropriate funnel category and compared to each of the 14 home voids performed by each subject using the funnel device where 0 = same category as office reading, +1 = home reading 1 category higher than office reading, +2 = home reading 2 categories higher than office reading, −1 = home reading 1 category lower than office reading, −2 = home reading 2 categories lower than office reading. The difference in voided volume was calculated as: home Voffice Vvoid - office Vvoid
| 1. | Place of publication | : | Mumbai |
| 2. | Periodicity of its publication | : | Quarterly (January, April, July and October) |
| 3. | Printer's Name | : | Medknow Publications Pvt. Ltd. |
| Nationality | : | Indian | |
| Address | : | A-108/109, Kanara Business Center, | |
| Off Link Rd, Ghatkopar (E), | |||
| Mumbai - 400075, India | |||
| Phone: 91-22-6649 1818/1816, | |||
| Fax: 91-22-6649 1817 | |||
| 4. | Publisher's Name | : | Medknow Publications Pvt. Ltd. |
| Nationality | : | Indian | |
| Address | : | A-108/109, Kanara Business Center, | |
| Off Link Rd, Ghatkopar (E), | |||
| Mumbai - 400075, India | |||
| Phone: 91-22-6649 1818/1816, | |||
| Fax: 91-22-6649 1817 | |||
| 5. | Editor's Name | : | Nitin S. Kekre |
| Nationality | : | Indian | |
| Address | : | Department of Urology, Christian Medical College, | |
| Vellore - 632 004, Tamil Nadu, India | |||
| Phone: +91-0416-2282111, Fax: +91- 0416 - 2232035, | |||
| E-mail: editor@indianjurol.com. | |||
| 6. | Names and addresses of individuals who own the newspaper and partners or shareholders holding more than one per cent of the total capital. | : | Urological Society of India |
| I, | |||
| Date: 12th April, 2007 | Editor | ||