| Literature DB >> 19670869 |
Qiang Zhao1, Xiufen Lu, Chun-Gang Yuan, Xing-Fang Li, X Chris Le.
Abstract
We describe a sensitive and specific sandwich assay for human alpha-thrombin. The assay takes advantage of sandwich binding of two affinity aptamers for increased specificity, gold nanoparticles for signal amplification, magnetic beads for fast magnetic separation, and inductively coupled plasma mass spectrometry for ultrasensitive detection. Other proteins, such as immunoglobulin G, serum albumin, transferrin, fibrinogen, and lysozyme did not show interference with the assay for human alpha-thrombin. The detection limit of human alpha-thrombin was as low as 0.5 fmol, corresponding to 10 pM thrombin in 50 microL, and the dynamic range covered approximately 3 orders of magnitude.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19670869 DOI: 10.1021/ac900961y
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986