Literature DB >> 19659602

Comparison of the Cobalt Glidescope video laryngoscope with conventional laryngoscopy in simulated normal and difficult infant airways.

Michelle White1, Nicola Weale, Judith Nolan, Steven Sale, Guy Bayley.   

Abstract

AIM: To evaluate the new pediatric Glidescope (Cobalt GVL Stat) by assessing the time taken to tracheal intubation under normal and difficult intubation conditions. We hypothesized that the Glidescope would perform as well as conventional laryngoscopy.
BACKGROUND: A new pediatric Glidescope became available in October 2008. It combines a disposable, sterile laryngoscope blade and a reusable video baton. It is narrower and longer than the previous version and is available in a greater range of sizes more appropriate to pediatric use.
METHODS: We performed a randomized study of 32 pediatric anesthetists and intensivists to compare the Cobalt GVL Stat with the Miller laryngoscope under simulated normal and difficult airway conditions in a pediatric manikin.
RESULTS: We found no difference in time taken to tracheal intubation using the Glidescope or Miller laryngoscope under normal (29.3 vs 26.2 s, P = 0.36) or difficult (45.8 and 44.4 s, P = 0.84) conditions. Subjective evaluation of devices for field of view (excellent: 59% vs 53%) and ease of use (excellent: 69% vs 63%) was similar for the Miller laryngoscope and Glidescope, respectively. However, only 34% of participants said that they would definitely use the Glidescope in an emergency compared with 66% who would be willing to use the Miller laryngoscope.
CONCLUSIONS: The new Glidescope performs as well as the Miller laryngoscope under simulated normal and difficult airway conditions.

Entities:  

Mesh:

Year:  2009        PMID: 19659602     DOI: 10.1111/j.1460-9592.2009.03123.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  5 in total

1.  A comparison of two hyperangulated video laryngoscope blades to direct laryngoscopy in a simulated infant airway: a bicentric, comparative, randomized manikin study.

Authors:  Marc Kriege; Nina Pirlich; Thomas Ott; Eva Wittenmeier; Frank Dette
Journal:  BMC Anesthesiol       Date:  2018-08-31       Impact factor: 2.217

2.  GlideScope® cobalt video laryngoscope versus direct Miller laryngoscope for lateral position-tracheal intubation in neonates with myelodysplasia: A prospective randomized study.

Authors:  Eman Ramadan Salama; Doaa El Amrousy
Journal:  Saudi J Anaesth       Date:  2019 Jan-Mar

3.  Comparison of GlideScope Videolaryngoscopy to Direct Laryngoscopy for Intubation of a Pediatric Simulator by Novice Physicians.

Authors:  Joni E Rabiner; Marc Auerbach; Jeffrey R Avner; Dina Daswani; Hnin Khine
Journal:  Emerg Med Int       Date:  2013-10-31       Impact factor: 1.112

4.  Novel technique for placement of laryngeal mask airway in difficult pediatric airways.

Authors:  Fatemeh Roodneshin; Mahvash Agah
Journal:  Tanaffos       Date:  2011

5.  Pediatric Trainees Managing a Difficult Airway: Comparison of Laryngeal Mask Airway, Direct, and Video-Assisted Laryngoscopy.

Authors:  Art Ambrosio; Kastley Marvin; Colleen Perez; Chelsie Byrnes; Cory Gaconnet; Chris Cornelissen; Matthew Brigger
Journal:  OTO Open       Date:  2017-05-08