Literature DB >> 19651274

The sensitivity and specificity of an activity monitor in detecting functional activities in young people with cerebral palsy.

Anna H Mackey1, Penelope Hewart, Sharon E Walt, N Susan Stott.   

Abstract

OBJECTIVE: To evaluate the sensitivity and specificity of the Intelligent Device for Energy Expenditure and Activity (IDEEA) monitor in detecting functional activities in young people with cerebral palsy (CP).
DESIGN: Validation study.
SETTING: Two identical data collections completed 1 week apart at a gait laboratory. PARTICIPANTS: Twenty-five young people with CP and 30 able-bodied (AB) peers (age 8-25 y; mean age CP 14.1 y, Gross Motor Function Classification System Level I-III; mean age AB, 14.2 y).
INTERVENTIONS: Not applicable. MAIN OUTCOME MEASURES: Each participant completed 5 functional activities (sitting, lying, standing, walking, and stairs), with the monitor attached. The sensitivity and specificity of the monitor in detecting each activity was calculated by comparison to a written timed report.
RESULTS: Sitting, lying, and standing were detected with median 100% sensitivity in both participant groups and across both testing sessions. Accuracy of walking detection was reduced compared with static activities (P<.05) across the 2 sessions and groups. The CP group had a significantly higher number of participants where the activity was not detected with 100% sensitivity (lie: 18% CP, 5% AB, P<.04; stand: 12% CP, 0% AB, P<.02; walk: 96% CP, 81% AB, P<.03). Stair climbing was detected in only half of the 12 participants with CP who could achieve the task. The IDEEA demonstrated high specificity (range, 97%-100%) for both participant groups.
CONCLUSIONS: The IDEEA activity monitor had high levels of sensitivity and specificity in determining everyday static activities in participants with CP, with greater difficulties in detecting dynamic activities of walking and stair climbing.

Entities:  

Mesh:

Year:  2009        PMID: 19651274     DOI: 10.1016/j.apmr.2009.01.029

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  2 in total

1.  [Instrumented gait and movement analysis of musculoskeletal diseases].

Authors:  K Sander; D Rosenbaum; H Böhm; F Layher; T Lindner; R Wegener; S I Wolf; F Seehaus
Journal:  Orthopade       Date:  2012-10       Impact factor: 1.087

2.  IDEEA activity monitor: validity of activity recognition for lying, reclining, sitting and standing.

Authors:  Yuyu Jiang; Janet L Larson
Journal:  Front Med       Date:  2012-12-22       Impact factor: 4.592

  2 in total