A device and procedure for immobilization of patients receiving limb-preserving radiotherapy for soft tissue sarcoma.

Colleen I Dickie1, Amy Parent, Anthony Griffin, Tim Craig, Charles Catton, Peter Chung, Tony Panzarella, Brian O'Sullivan, Michael Sharpe.   

Abstract

The purpose of this study was to determine the accuracy and efficiency of a custom-designed immobilization device for patients with extremity soft-tissue sarcoma. The custom device consisted of a thermoplastic shell, vacuum pillow, and adaptable baseplate. The study included patients treated from January 2005 to March 2007, with 92 patients immobilized with the custom device and 98 with an established standard. Setup times for these cohorts were analyzed retrospectively for conformal and intensity modulated radiotherapy techniques (IMRT). Thigh tumor setup times were analyzed independently. A subset of patients treated with IMRT was analyzed for setup error using the radiographically verified isocenter position measured daily with electronic portal imaging and cone-beam computed tomography. Mean setup time was reduced by 2.2 minutes when using the custom device for conformal treatment (p = 0.03) and by 5.8 min for IMRT of thigh tumors (p = 0.009). All other setup time comparisons were not significant. A significant systematic error reduction was seen in all directions using the custom device. Random error standard deviations favored the custom device. The custom device offers immobilization advantages. Patient setup time was reduced for conformal techniques and IMRT of thigh tumors. Positioning uncertainty was improved, permitting a reduction of the planning target volume margin by 2 to 4 mm.

Entities:  

Mesh:

Year:  2008        PMID: 19647636     DOI: 10.1016/j.meddos.2008.10.003

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  2 in total

1.  Long-time clinical experience in patient setup for several particle therapy clinical indications: management of patient positioning and evaluation of setup reproducibility and stability.

Authors:  Rosalinda Ricotti; Andrea Pella; Barbara Tagaste; Giovanni Elisei; Giulia Fontana; Maria Bonora; Mario Ciocca; Francesca Valvo; Roberto Orecchia; Guido Baroni
Journal:  Br J Radiol       Date:  2019-11-14       Impact factor: 3.629

2.  Evidence-based region of interest matching guidelines for sarcoma volumetric image-guided radiation therapy.

Authors:  Winnie Li; Samuel Appiah; Christine Hill; Nathan Becker; Charles Catton; Peter Chung; David Shultz; Peter Ferguson; Brian O'Sullivan; Colleen Dickie
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2018-02-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.