Literature DB >> 19646318

Accuracy of bladder scanning in the assessment of postvoid residual volume.

Ghadeer Al-Shaikh1, Annick Larochelle2, Craig E Campbell1, Joyce Schachter1, Kevin Baker1, Dante Pascali1.   

Abstract

OBJECTIVE: To compare the accuracy of the 3D portable ultrasound with catheterization in the assessment of postvoid residual (PVR) urine volume among women in the urogynaecology clinic.
METHODS: A prospective study was performed, assessing 101 women. After the patient voided four ultrasound (US) assessments were carried out using the BladderScan BVI 3000; the patient was then catheterized. The reproducibility of the US measurements and the difference between the two methods were assessed using Bland and Altman plots. The strength of the relationship was measured by a simple Pearson correlation coefficient.
RESULTS: The results showed that 3D scanner measurements were highly reproducible and were also found to correlate significantly with catheterized volume (r=0.79, 95% CI 0.70-0.85, P<0.001). The mean difference between the two methods was 12.9 mL (95% CI 5.5-20.2 mL, P<0.001).
CONCLUSION: In determining PVR volumes, the portable ultrasound BladderScan BVI 3000 is an accurate alternative to bladder catheterization.

Entities:  

Mesh:

Year:  2009        PMID: 19646318     DOI: 10.1016/S1701-2163(16)34215-3

Source DB:  PubMed          Journal:  J Obstet Gynaecol Can        ISSN: 1701-2163


  9 in total

1.  Accuracy and precision of a new portable ultrasound scanner, the Biocon-700, in residual urine volume measurement.

Authors:  Moon Kyoung Cho; Eun Ji Noh; Chul Hong Kim
Journal:  Int Urogynecol J       Date:  2016-12-10       Impact factor: 2.894

2.  Repeatability of post-void residual urine ≥ 100 ml in urogynaecologic patients.

Authors:  Marie-Louise Saaby; Gunnar Lose
Journal:  Int Urogynecol J       Date:  2011-12-13       Impact factor: 2.894

3.  The accuracy of three-dimensional bladder ultrasonography in determining the residual urinary volume compared with conventional catheterisation.

Authors:  Imran K Jalbani; M Hammad Ather
Journal:  Arab J Urol       Date:  2014-07-29

4.  Is bedside ultrasound now a routine part of specialist obstetrics and gynaecology outpatient consultation?

Authors:  Sarah Van Der Wal; Stephen J Robson; Shawn Choong
Journal:  Australas J Ultrasound Med       Date:  2015-12-31

5.  Falsely Elevated Postvoid Residual Urine Volume in Uterine Myoma.

Authors:  Tae Hee Kim; Hyo Sang Kim; Jung Wook Park; Oh Kyung Lim; Ki Deok Park; Ju Kang Lee
Journal:  Ann Rehabil Med       Date:  2017-04-27

6.  False positive bladder scan in ascites with anuria.

Authors:  Sarosh Janardanan; Ahmed E M Moussa; Philip James
Journal:  Clin Case Rep       Date:  2019-07-03

7.  The Accuracy of Portable Ultrasound Bladder Scanner Measurements of Postvoid Residual Volume in Women With Pelvic Organ Prolapse.

Authors:  John Graham Theisen; Nicolette E Deveneau; Anu Agrawal; Casey Kinman; Jeremy Gaskins; Kate Meriwether; Sean L Francis
Journal:  Female Pelvic Med Reconstr Surg       Date:  2019 Sep/Oct       Impact factor: 2.091

Review 8.  Prevention and management of postoperative urinary retention after urogynecologic surgery.

Authors:  Elizabeth J Geller
Journal:  Int J Womens Health       Date:  2014-08-28

9.  Circadian Patterns in Postvoid Residual and Voided Percentage Among Older Women with Urinary Incontinence.

Authors:  Veerle Decalf; Thomas F Monaghan; Marie-Astrid Denys; Mirko Petrovic; Ronny Pieters; Jeffrey P Weiss; Karel Everaert
Journal:  J Clin Med       Date:  2020-03-27       Impact factor: 4.241

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.