Factors that influence measurements and accurate evaluation of stent apposition by optical coherence tomography. Assessment using a phantom model.

Takahiro Sawada1, Junya Shite, Noriyuki Negi, Toshiro Shinke, Yusuke Tanino, Daisuke Ogasawara, Hiroyuki Kawamori, Hiroki Kato, Naoki Miyoshi, Naoki Yoshino, Amane Kozuki, Masanobu Koto, Ken-ichi Hirata.   

Abstract

BACKGROUND: Factors influencing measurements and accurate evaluation of stent apposition by optical coherence tomography (OCT) are not established. METHODS AND
RESULTS: Phantom models of known luminal sizes and lengths were evaluated by OCT under various conditions and measurements were compared with actual values. Stents implanted into phantom models were examined by OCT to validate the measurement point on the stent strut surface for accurate evaluation of stent apposition. Strut thickness was measured at 3 points (midpoint, inner and outer surfaces of the stent shadow). The precision of OCT measurements of lumen diameter and area was satisfactory when the image wire was positioned in the center, but the error and deviation were unsatisfactory when the image wire was positioned eccentrically using a low frame acquisition rate. Longitudinal OCT measurements were close to actual values under all conditions examined. Measurements from the midpoint of the stent shadow to the adjacent vessel wall surface coincided with actual stent thickness.
CONCLUSIONS: Significant measurement error can occur if the image wire is positioned eccentrically with a lower frame acquisition rate than specified by the manufacturer. To accurately evaluate stent apposition, the stent surface should be measured from the center of the stent reflection.

Mesh:

Year:  2009        PMID: 19644218     DOI: 10.1253/circj.cj-09-0113

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  8 in total

1.  Wall Apposition Is a Key Factor for Aneurysm Occlusion after Flow Diversion: A Histologic Evaluation in 41 Rabbits.

Authors:  A Rouchaud; C Ramana; W Brinjikji; Y-H Ding; D Dai; T Gunderson; J Cebral; D F Kallmes; R Kadirvel
Journal:  AJNR Am J Neuroradiol       Date:  2016-07-07       Impact factor: 3.825

2.  Automatic segmentation of in-vivo intra-coronary optical coherence tomography images to assess stent strut apposition and coverage.

Authors:  G J Ughi; T Adriaenssens; K Onsea; P Kayaert; C Dubois; P Sinnaeve; M Coosemans; W Desmet; J D'hooge
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-24       Impact factor: 2.357

3.  Fusion of 3D QCA and IVUS/OCT.

Authors:  Shengxian Tu; Niels R Holm; Gerhard Koning; Zheng Huang; Johan H C Reiber
Journal:  Int J Cardiovasc Imaging       Date:  2011-01-25       Impact factor: 2.357

Review 4.  The role of optical coherence tomography in coronary intervention.

Authors:  Mitsuyasu Terashima; Hideaki Kaneda; Takahiko Suzuki
Journal:  Korean J Intern Med       Date:  2012-02-28       Impact factor: 2.884

5.  Effect of proximal balloon edge dilation technique for opening a side branch ostium in repetitive-proximal optimizing technique sequence.

Authors:  Teruyoshi Kume; Yoshinobu Murasato; Ryotaro Yamada; Terumasa Koyama; Tomoko Tamada; Koichiro Imai; Hiroshi Okamoto; Ayano Enzan; Yoji Neishi; Shiro Uemura
Journal:  Catheter Cardiovasc Interv       Date:  2020-04-24       Impact factor: 2.692

6.  Vascular Healing Response after Everolimus-Eluting Stent Implantation in Acute Coronary Syndrome Culprit Lesions: Comparison with Implantation in Stable Angina Pectoris.

Authors:  Yutaka Goryo; Teruyoshi Kume; Tomoya Ueda; Makoto Watanabe; Ryotaro Yamada; Yoji Neishi; Yoshihiko Saito; Shiro Uemura
Journal:  Acta Cardiol Sin       Date:  2018-03       Impact factor: 2.672

Review 7.  Research and clinical applications of optical coherence tomography in invasive cardiology: a review.

Authors:  Luigi Vignali; Emilia Solinas; Enzo Emanuele
Journal:  Curr Cardiol Rev       Date:  2014-11

8.  Influence of dual antiplatelet therapy duration on neointimal condition after second-generation drug-eluting stent implantation.

Authors:  Yutaka Goryo; Teruyoshi Kume; Hiroshi Okamoto; Ai Kawamura; Kenzo Fukuhara; Tomoko Tamada; Terumasa Koyama; Koichiro Imai; Ryotaro Yamada; Yoji Neishi; Shiro Uemura
Journal:  Cardiovasc Interv Ther       Date:  2021-02-26
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.