Identification of performance problems in a commercial human immunodeficiency virus type 1 enzyme immunoassay by multiuser external quality control monitoring and real-time data analysis.

J Kim1, C Swantee, B Lee, H Gunning, A Chow, F Sidaway, C Sherlock, R Garceau, W Dimech, L Malloch.   

Abstract

In June 2005, a pilot program was implemented in Canadian laboratories to monitor the performance of the Abbott human immunodeficiency virus types 1 and 2 (HIV-1/2) gO enzyme immunoassay (EIA). Two different external quality control (QC) reagents and a "real-time" software analysis program were evaluated. In November 2005, higher-than-expected calibrator rate values in these kits were first reported at the Ontario Ministry of Health (Etobicoke), followed by the Alberta Provincial Public Health Laboratory (Edmonton and Calgary) and others. These aberrations were easily and readily tracked in "real time" using the external QC reagents and the software program. These high calibrator values were confirmed in Delkenheim, Germany, by Abbott, and a manufacturing change was initiated beginning with lot 38299LU00, which was distributed to laboratories in Canada in April 2006. However, widespread reports of calibrator failure by laboratories outside Canada were made in March 2006. In April 2006, Abbott Diagnostics initiated a level III investigation to identify the root cause, which was prolonged storage, under uncontrolled storage conditions, of the raw material used in the manufacture of the matrix cells. To the best of our knowledge, this is the first example of a program in Canada for serological testing that combines a common external QC reagent and a "real-time" software program to allow laboratories to monitor kit performance. In this case, external QC monitoring helped identify and confirm performance problems in the Abbott HIV-1/2 gO EIA kit, further highlighting the benefit of implementing such a program in a national or multilaboratory setting for laboratories performing diagnostic and clinical monitoring testing.

Entities:  

Mesh:

Year:  2009        PMID: 19641055      PMCID: PMC2756918          DOI: 10.1128/JCM.00892-09

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  3 in total

1.  Improvement in the performance of HIV screening kits.

Authors:  K R Perry; S Ramskill; R P Eglin; J A J Barbara; J V Parry
Journal:  Transfus Med       Date:  2008-08       Impact factor: 2.019

2.  Dynamics of HIV viremia and antibody seroconversion in plasma donors: implications for diagnosis and staging of primary HIV infection.

Authors:  Eberhard W Fiebig; David J Wright; Bhupat D Rawal; Patricia E Garrett; Richard T Schumacher; Lorraine Peddada; Charles Heldebrant; Richard Smith; Andrew Conrad; Steven H Kleinman; Michael P Busch
Journal:  AIDS       Date:  2003-09-05       Impact factor: 4.177

3.  Sensitivity of HIV-antibody assays determined by seroconversion panels.

Authors:  N T Constantine; G van der Groen; E M Belsey; H Tamashiro
Journal:  AIDS       Date:  1994-12       Impact factor: 4.177

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.