Literature DB >> 19641009

Design, implementation and results of the quality control program for the Australian government's point of care testing in general practice trial.

Mark Shephard1, Anne Shephard, Les Watkinson, Beryl Mazzachi, Paul Worley.   

Abstract

BACKGROUND: From 2005 to 2007 the Australian Government funded a multicentre, clustered randomized controlled trial to determine the clinical effectiveness, cost-effectiveness, satisfaction and safety of point of care testing (PoCT) in general practice (GP). PoC tests measured (and devices used) in the trial were haemoglobin A1c and urine albumin:creatinine ratio (DCA 2000), lipids (Cholestech LDX) and international normalized ratio (CoaguChek S).
METHODS: An internal quality control (QC) program was developed as part of a quality management framework for the trial. PoCT device operators were provided with a colour-coded QC Result Sheet and QC Action Sheet for on-site recording and interpreting of their results. Within-practice imprecision for QC testing was calculated and compared with the analytical goals for imprecision set prior to the trial.
RESULTS: The average participation rate for QC testing was 91% or greater. Median within-practice imprecision met the analytical goals for all PoC tests, except for high-density lipoprotein-cholesterol (HDL-C) where observed performance was outside the minimum goal for one level and one lot number of QC. Most practices achieved the imprecision goals for all analytes, with the principal exception of HDL-C.
CONCLUSIONS: Results from QC testing indicate that PoCT in the GP trial met the analytical goals set for the trial, with the exception of HDL-C.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19641009     DOI: 10.1258/acb.2009.009045

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  5 in total

1.  The Conduct of Quality Control and Quality Assurance Testing for PoCT Outside the Laboratory.

Authors:  Janice P Gill; Mark Ds Shephard
Journal:  Clin Biochem Rev       Date:  2010-08

2.  The National QAAMS Program - A Practical Example of PoCT Working in the Community.

Authors:  Mark Ds Shephard; Janice P Gill
Journal:  Clin Biochem Rev       Date:  2010-08

Review 3.  Impact of HbA1c Testing at Point of Care on Diabetes Management.

Authors:  Oliver Schnell; J Benjamin Crocker; Jianping Weng
Journal:  J Diabetes Sci Technol       Date:  2016-11-27

4.  Educational intervention to improve effectiveness in treatment and control of patients with high cardiovascular risk in low-resource settings in Argentina: study protocol of a cluster randomised controlled trial.

Authors:  Pablo Gulayin; Vilma Irazola; Alfredo Lozada; Martin Chaparro; Marilina Santero; Laura Gutierrez; Rosana Poggio; Andrea Beratarrechea; Adolfo Rubinstein
Journal:  BMJ Open       Date:  2017-01-31       Impact factor: 2.692

5.  Results of a survey among GP practices on how they manage patient safety aspects related to point-of-care testing in every day practice.

Authors:  Claudette de Vries; Carine Doggen; Ellen Hilbers; Robert Verheij; Maarten IJzerman; Robert Geertsma; Ron Kusters
Journal:  BMC Fam Pract       Date:  2015-02-05       Impact factor: 2.497

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.